Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
129
Additional information
Selection
For a detailed description of the options available, see the
Measuring mode
Examples
For a detailed description of the configuration examples, see the
Measuring mode
Damping out. 1 to n
Navigation
Expert → Output → PFS output 1 to n → Damping out. 1 to n (0477–1 to n)
Prerequisite
One of the following options is selected in the
Assign curr.
• Volume flow
• Mass flow
• Correct.vol.flow
• Flow velocity
• Conductivity
• CorrConductivity
• Temperature
• Electronic temp.
Description
Use this function to enter a time constant for the reaction time of the output signal to
fluctuations in the measured value.
User entry
0 to 999.9 s
Factory setting
0.0 s
Additional information
User entry
Use this function to enter a time constant (PT1 element
) for frequency output damping:
• If a low time constant is entered, the current output reacts particularly quickly to
fluctuating measured variables.
• On the other hand, the current output reacts more slowly if a high time constant is
entered.
Damping is switched off if
0
is entered (factory setting).
The frequency output is subject to separate damping that is independent of all preceding
time constants.
*
Visibility depends on order options or device settings
4)
proportional transmission behavior with first order delay