4 Wiring
PROline Promag 23
36
Hauser
4.2.2
Special cases
Metal, ungrounded piping
In order to prevent outside influences on measurement, it is advisable to use ground
cables to connect each sensor flange to its corresponding pipe flange and ground the
flanges. Connect the transmitter or sensor connection housing, as applicable, to ground
potential by means of the ground terminal provided for the purpose (Fig. 25).
"
Caution:
Please also pay special attention to company-internal grounding concepts.
!
Note:
The ground cable for flange-to-flange connections can be ordered separately as an
accessory from E+H (see Page 59). The ground cable is in direct connection with the
conductive flange coating and is secured by the flange screws.
Fig. 25:
Potential equalisation with equalising currents in metallic, non-grounded piping systems
6 mm² Cu
F06-2xxxxx
xx-04-xx
-xx-x
x
-001