Proline Promag 10D
Hauser
11
Installation instructions
Installation with pumps
The sensor should only be installed on the pump pressure side.
!
Note!
• The sensor should
never
be installed on the pump suction side in order to avoid the risk of low pressure,
and thus damage to the measuring tube.
Information on the pressure tightness of the measuring tube
ä
15, section "Pressure tightness".
• Pulsation dampers may be needed if the sensor is installed downstream from piston pumps, piston diaphragm
pumps or hose pumps.
Information on the shock and vibration resistance of the device
ä
15, section "Shock and vibration
resistance".
A0010706
Installing the device with a pump
Partially filled pipes
Partially filled pipes with gradients necessitate a drain-type configuration.
A0010707
Installation with partially filled pipes
Down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes h
5 m (16,4 ft). This precaution is
to avoid low pressure and the consequent risk of damage to the measuring tube. This measure also prevents
the system losing prime, which could cause air pockets. Information on the pressure tightness of the measuring
tube
ä
15, section "Pressure tightness".
A0010708
Installation measures for down pipes
1
Vent valve
2
Pipe siphon
h
Length of the down pipe
³
5 x DN
³
2 x DN
h
2
1