Device functions - vortex measuring devices
Proline Fieldcheck
84
Hauser
Function group OPERATION
Function description
FUNCTION
SIMULATION
OPERATION
SIMULATE
Use this function to control the simulation.
Options:
START
STOP
Factory setting:
START
Display during the simulation:
A0013496
Fig. 30:
Example of information displayed during simulation
Line 1 = Function: Simulation
Line 2 = The flow value generated by Fieldcheck with the Simubox
Line 3 = The measured value returned by the measuring device
Line 4 = Empty
Line 5 = Current, frequency, etc.
In the SIMULATION operating mode, the current values of the current and frequency
output of the tested device are displayed (multimeter function). Pulses are displayed as
frequency if there is more than 1 pulse/second at the frequency input of Fieldcheck.
In the STOP status the simulation signal value 0 is always output.
Affected settings for the simulation, however, are maintained.
Simulation
1500.0
dm3/m
1501.3
dm3/m
I=12.003 mA,
F=229 Hz