Proline Cubemass C 100 HART
Electrical connection
Hauser
33
5. To ensure that moisture does not enter the cable entry:
Route the cable so that it loops down before the cable entry ("water trap").
A0029278
6. Insert dummy plugs into unused cable entries.
7.6
Post-connection check
Are cables or the device undamaged (visual inspection)?
Do the cables used meet the requirements → 26?
Do the cables have adequate strain relief?
Are all the cable glands installed, firmly tightened and leak-tight? Cable run with "water trap"
→ 32?
Depending on the device version: are all the device plugs firmly tightened → 29?
Does the supply voltage match the specifications on the transmitter nameplate → 105?
Is the terminal assignment → 27 or pin assignment of the connector → 28 correct?
If supply voltage is present, is the power LED on the electronics module of the transmitter lit green
→ 11?
Depending on the device version, is the securing clamp or fixing screw firmly tightened?