Electrical connection
Proline 800 Cellular radio
36
Hauser
PE = P
P
=
P
M
A0044854
‣
Connect the connection housing of the transmitter or sensor to ground potential via the
ground terminal provided for this purpose.
Metal pipe without liner
• Potential equalization is via the ground terminal and pipe flanges.
• The medium is set to ground potential.
Starting conditions:
• Pipes are not sufficiently grounded.
• Pipes are conductive and at the same electrical potential as the medium
DN
300
≤
DN
350
≥
PE = P
P
=
P
M
A0042089
1. Connect both sensor flanges to the pipe flange via a ground cable and ground them.