background image

Products

Solutions

Services

Brief Operating Instructions

Proline 800

with electromagnetic sensor

Cellular radio (Type: DA800C-CRM02.01)

These instructions are Brief Operating Instructions; they are

not

 a substitute for the Operating Instructions pertaining to

the device.

Brief Operating Instructions part 2 of 2: Transmitter

Contain information about the transmitter.
Brief Operating Instructions part 1 of 2: Sensor →   3

KA01495D/06/EN/02.22-00

71552846

2022-01-15

Содержание Proline 800

Страница 1: ...2 01 These instructions are Brief Operating Instructions they are not a substitute for the Operating Instructions pertaining to the device Brief Operating Instructions part 2 of 2 Transmitter Contain...

Страница 2: ...Proline 800 Cellular radio 2 Endress Hauser Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555...

Страница 3: ...ing device Incoming acceptance and product identification Storage and transport Installation Brief Operating Instructions Part 2 Transmitter The Transmitter Brief Operating Instructions are aimed at s...

Страница 4: ...3 Connecting the measuring device 22 5 4 Power supply via battery packs Proline 800 Advanced 26 5 5 Connect the cellular antenna 29 5 6 Connecting the pressure sensor Proline 800 Advanced 30 5 7 Powe...

Страница 5: ...ymbols Symbol Meaning Symbol Meaning Direct current Alternating current Direct current and alternating current Ground connection A grounded terminal which as far as the operator is concerned is ground...

Страница 6: ...esses or actions that are permitted Preferred Procedures processes or actions that are preferred Forbidden Procedures processes or actions that are forbidden Tip Indicates additional information Refer...

Страница 7: ...ithin the specified pressure and temperature range Only use the measuring device in full compliance with the data on the nameplate and the general conditions listed in the Operating Instructions and s...

Страница 8: ...t via the measuring device If working on and with the device with wet hands Due to the increased risk of electric shock wear suitable gloves 2 4 Operational safety Risk of injury Operate the device in...

Страница 9: ...to support protective measures on the operator s side These functions can be configured by the user and guarantee greater in operation safety if used correctly For detailed information on device speci...

Страница 10: ...ted maximum ambient temperature If operating outdoors Avoid direct sunlight and exposure to weathering particularly in warm climatic regions LCAUTION Excessive force can damage the housing Avoid exces...

Страница 11: ...sk of damaging the plastic transmitter Tighten the fixing screws as per the tightening torque 2 Nm 1 5 lbf ft 20 70 0 79 2 75 102 4 0 4 x SW 10 3 x TX 25 A0029051 2 Engineering unit mm in 4 1 3 Turnin...

Страница 12: ...20 A0044272 1 Loosen the fixing screws of the housing cover when reassembling pay attention to the tightening torque 15 2 Open the housing cover ON ON 1 2 3 4 5 A0044274 3 Open the display module 4 Pu...

Страница 13: ...ser 13 6 7 A0044273 6 Push the display module down gently at the hinge 7 Pull the display module out of the holder A0043338 8 Disconnect the plug of the electronics module T10 1 Nm 0 7 lbf ft A0042853...

Страница 14: ...10 Remove the electronics module P U S H T O R E M O V E TX 20 4 x 11 P U S H T O R E M O V E 12 A0044276 11 Loosen the fixing screws of the main electronics module when reassembling pay attention to...

Страница 15: ...he desired position in increments of 90 Reassembling the transmitter housing LWARNING Excessive tightening torque applied to the fixing screws Risk of damaging the plastic transmitter Tighten the fixi...

Страница 16: ...6 Endress Hauser P U S H T O R E M O V E A0044279 Reverse the procedure to reassemble the measuring device 4 1 4 Mounting the external cellular antenna If the signal is too weak mount the external cel...

Страница 17: ...r mounting the antenna 15 Nm 2 Antenna horizontal installation maximum torque for mounting the antenna 15 Nm 3 Grounding lug with Torx screw and washer 4 Torx screw with washer 5 Antenna holder 4 1 5...

Страница 18: ...ard that has been inserted The additionally inserted SIM card may not be protected by a PIN Only 1 8 V SIM cards are recognized 1 2 1 2 T20 A0042838 A0044342 Open the connection compartment cover Remo...

Страница 19: ...s Mounting the transmitter housing Post mounting Wall mounting Turning the transmitter housing Is the device undamaged visual inspection Turning the transmitter housing Is the securing screw firmly ti...

Страница 20: ...or cable entries use appropriate tool Wire stripper When using stranded cables crimper for wire end ferrule For removing cables from terminal flat blade screwdriver 3 mm 0 12 in 5 2 2 Requirements for...

Страница 21: ...6 to 12 mm 0 24 to 0 47 in For reinforced cable M20 1 5 with cable 9 5 to 16 mm 0 37 to 0 63 in Plug in spring terminals for wire cross sections 0 5 to 2 5 mm2 20 to 14 AWG 5 2 3 Terminal assignment...

Страница 22: ...Observe applicable federal national installation codes and regulations Comply with local workplace safety regulations Observe grounding concept of the plant Never mount or wire the measuring device w...

Страница 23: ...sh the cable through the cable entry To ensure tight sealing do not remove the sealing ring from the cable entry 4 Strip the cable and cable ends In the case of stranded cables also fit ferrules Verwe...

Страница 24: ...cable ends In the case of stranded cables also fit ferrules Verweisziel existiert nicht aber y link required true 5 Connect the cable in accordance with the terminal assignment 6 Firmly tighten the ca...

Страница 25: ...1 Loosen the 4 fixing screws on the housing cover 2 Open the housing cover 3 Push the cable through the cable entry To ensure tight sealing do not remove the sealing ring from the cable entry 4 Strip...

Страница 26: ...f protection may be voided due to insufficient sealing of the housing Screw in the screw without using any lubricant Tighten the 4 fixing screws on the housing cover 5 4 Power supply via battery packs...

Страница 27: ...ate the device it is important to ensure that switch B is set to ON and the ribbon cable is connected to the electronics module The device starts once the buffer capacitor is connected After 15 second...

Страница 28: ...15 seconds a measured value appears on the display 8 9 A0043732 A0043733 Insert battery packs 1 and 2 Plug the connector of battery pack 1 into connector 1 Plug the connector of battery pack 2 into co...

Страница 29: ...unt the external cellular antenna in a separate location 1 2 1 2 T20 1 2 A0042838 A0044338 Open the connection compartment cover Plug in the antenna connector 3 A B A0044324 Guide the cable of the cel...

Страница 30: ...Close the connection compartment cover 5 6 Connecting the pressure sensor Proline 800 Advanced A B A0044314 A Plug for the pressure sensor on the housing of the compact transmitter B Plug for the pre...

Страница 31: ...nection Endress Hauser 31 5 7 Power supply via the external battery pack Proline 800 Advanced 5 7 1 Connecting the external battery pack A A0044313 A Plug for the external battery pack Connect the ext...

Страница 32: ...ining capacity of the external batteries The value on the display refers exclusively to the batteries fitted internally If internal and external batteries are connected the external batteries are used...

Страница 33: ...ualization is not used Equalizing currents are present Ground cable Copper wire at least 6 mm2 0 0093 in2 DN 300 DN 350 A0029338 7 Potential equalization via ground terminal and pipe flanges 1 Connect...

Страница 34: ...tuations where Standard company potential equalization cannot be guaranteed Equalizing currents can be expected A0029339 8 Potential equalization via ground terminal and ground disks PE PFL PM 1 Conne...

Страница 35: ...azard The following requirements must be observed to ensure correct trouble free measurement The principle that the medium the sensor and the transmitter must be at the same electrical potential appli...

Страница 36: ...ded for this purpose Metal pipe without liner Potential equalization is via the ground terminal and pipe flanges The medium is set to ground potential Starting conditions Pipes are not sufficiently gr...

Страница 37: ...potential Starting conditions The pipe has an insulating effect Low impedance medium grounding close to the sensor is not guaranteed Equalizing currents through the medium cannot be ruled out PE P P P...

Страница 38: ...ght Cable run with water trap Only for remote version is the sensor connected to the right transmitter Check the serial number on the nameplate of the sensor and transmitter Does the supply voltage ma...

Страница 39: ...configured via the SmartBlue app In this case the connection is established via the Bluetooth wireless technology interface Supported functions Device selection in Live List and access to the device l...

Страница 40: ...ress Hauser Service 7 System integration For detailed information on system integration see the Operating Instructions for the device 3 Overview of device description files Current version data for th...

Страница 41: ...e configuration of the measuring device from unintentional modification after commissioning Write protection via access code 41 Write protection via write protection switch 42 8 4 1 Write protection v...

Страница 42: ...ccess code this allows write access to the entire operating menu to be locked The parameter values are now read only and cannot be edited any more The following parameters can always be modified even...

Страница 43: ...he fault 9 1 Diagnostic message Maintenance required Maintenance required Output signal is still valid Out of specification Device is being operated outside the technical specification limits e g outs...

Страница 44: ...www addresses endress com 71552846 71552846...

Отзывы: