Electrical connection
Levelflex FMP56, FMP57 PROFIBUS PA
16
Hauser
1. Pull out the display module with a gentle rotational movement.
2. Turn the display module to the desired position: Max. 8 × 45 ° in each direction.
3. Feed the coiled cable into the gap between the housing and main electronics module
and plug the display module into the electronics compartment until it engages.
Closing the cover of the electronics compartment
1.
2.
2.5 Nm
3 mm
A0021451
1. Screw down the cover of the electronics compartment.
2. Turn the securing clamp 90 ° in the clockwise direction and, using an Allen key (3 mm),
tighten the screw of the securing clamp on the electronics compartment cover with
2.5 Nm.
6
Electrical connection
6.1
Connecting requirements
6.1.1
Terminal assignment