MMP20 SONO Mobil HD2 (option E)
Page 18
Hauser
Menu item: "CUSTOMIZE":
Here, you have the option of performing your own material calibrations or of customizing existing
calibrations to suit your particular needs. Two options are available for this:
1-point calibration:
-
Adjusts the calibration curve to the selected point.
-
The slope is not changed.
-
Just one material sample is required.
2-point calibration:
-
Creation of a linear calibration between two measured
points
-
Two material samples with different moisture values are
required
1-point calibration:
This material calibration option merely adjusts (offsets) the configured calibration. As the slope is not
changed, it is important at the start to select a calibration curve that suits the material.
NOTE:
You require a sample of the material being measured to perform a 1-point material
calibration. The moisture value must be determined with another method (weighing and
drying, or similar) before the calibration.
Procedure:
At the start of the calibration, the calibration memory
(01
–
15) to be overwritten must be selected with the
"Up"
and "Down"
keys. Press the "Measure" key
to
accept the setting.
Then set the percentage moisture using the "Up"
and
"Down"
keys, and again press the "Measure" key
to
confirm your entries.