4.3 Lead sealing.
In order to secure the data-logger from tampering push the
sealing wire through the holes in the cover screws.
(Lead seal set, order No. RDL10X-PA)
4.4 Ambient conditions
Please keep away from strong magnetic fields
(see Technical data: Interference).
Surroundings to IP 65 ingress protection.
Ambient temperature range: -25 °C to +55 °C.
5. Electrical connection
5.1 Connection hints
Should there be a possibility of high energy transients when
operating with long cable lengths it is recommended that an
overvoltage protector be connected.
Screened signal cables are recommended when using the
following:
-
RTDs
-
Using the serial interface
Attention:
Power supply minus, GND connetion (pion 4) of the interface,
analogue input minus (terminal 1) and terminal 5 of the digital
input are internally connected. Always use barriers when
operating in series (e.g. galvanic separation of the serial
interface).
4. Installation
28
Содержание mini-log b
Страница 2: ......
Страница 69: ...13 Accessoires et consommables 67 ...
Страница 70: ...13 Accessoires et consommables 68 ...
Страница 71: ......