Commissioning
Micropilot M FMR 240 with Foundation Fieldbus
68
E Hauser
display (008)
The
distance
measured from the reference point to the product surface and the
level
calculated
with the aid of the empty adjustment are displayed. Check whether the values correspond to the
actual level or the actual distance. The following cases can occur:
• Distance correct
−
level correct
→
continue with the next function, "
check distance
"
(051)
• Distance correct
−
level incorrect
→
Check "
empty calibr.
"
(005)
• Distance incorrect
−
level incorrect
→
continue with the next function, "
check distance
"
(051)
Function "check distance" (051)
This function triggers the mapping of interference echoes. To do so, the measured distance must be
compared with the actual distance to the product surface. The following options are available for
selection:
Selection:
• distance = ok
• dist. too small
• dist. too big
•
dist. unknown
• manual
L00_FMR2xxxxx-14-00-06-en-010
distance = ok
• mapping is carried out up to the currently measured echo
• The range to be suppressed is suggested in the "
range of mapping
"
(052)
function
Anyway, it is wise to carry out a mapping even in this case.
dist. too small
• At the moment, an interference is being evaluated
• Therefore, a mapping is carried out including the presently measured echoes
• The range to be suppressed is suggested in the "
range of mapping
"
(052)
function
⇒
E HAUSER
E
+
–
⇒
E HAUSER
E
+
–