Operating menu
Micropilot FMR10
50
Hauser
User entry
0.0 to 12 m
Factory setting
12 m
Distance
Navigation
Setup → Distance
Description
Shows the distance D that is currently measured from the reference point (lower edge of
flange / last thread of sensor) to the level.
User interface
0.0 to 12 m
Level
Navigation
Setup → Level
Description
Displays the measured level L (before linearization). The unit is defined in the Distance
unit parameter (factory setting = m).
User interface
–99
999.9 to 200
000.0 m
Factory setting
0.0 m
Signal quality
Navigation
Setup → Signal quality
Description
Displays the signal quality of the level echo. Meaning of the display options - Strong: The
evaluated echo exceeds the threshold by at least 10 dB. - Medium: The evaluated echo
exceeds the threshold by at least 5 dB. - Weak: The evaluated echo exceeds the threshold
by less than 5 dB. - No signal: The device does not find a usable echo. The signal quality
indicated in this parameter always refers to the currently evaluated echo, either the level
echo or the tank bottom echo. In case of a lost echo (Signal quality = No signal) the device
generates the following error message: Diagnostic echo lost = Warning (factory setting) or
Alarm, if the other option has been selected in Diagnostic echo lost.
User interface
• Strong
• Medium
• Weak
• No signal