Memosens COL37E

Incoming acceptance and product identification

Hauser

7

3  

Incoming acceptance and product identification

3.1  

Incoming acceptance

1. Verify that the packaging is undamaged.

 Notify the supplier of any damage to the packaging.

Keep the damaged packaging until the issue has been resolved.

2. Verify that the contents are undamaged.

 Notify the supplier of any damage to the delivery contents.

Keep the damaged goods until the issue has been resolved.

3. Check that the delivery is complete and nothing is missing.

 Compare the shipping documents with your order.

4. Pack the product for storage and transportation in such a way that it is protected

against impact and moisture.
 The original packaging offers the best protection.

Make sure to comply with the permitted ambient conditions.

If you have any questions, please contact your supplier or your local Sales Center.

3.2  

Product identification

3.2.1  

Nameplate

The following information about the device appears on the nameplate:

• Manufacturer identification

• Order ID

• Extended order code

• Serial number

Compare the information on the nameplate with the order.

3.2.2  

Product identification

Product page

www.endress.com/col37e

Interpreting the order code

The order code and serial number of your product can be found in the following locations:

• On the nameplate

• In the delivery papers

Obtaining information on the product

1. Open 

www.endress.com

.

2. Call up the site search (magnifying glass).
3. Enter a valid serial number.

Содержание Memosens COL37E

Страница 1: ...Services Operating Instructions Memosens COL37E Agile optical oxygen sensor for laboratory measurements and random sampling in the field Digital with Memosens 2 0 technology BA02236C 07 EN 01 21 00 7...

Страница 2: ......

Страница 3: ...ng acceptance 7 3 2 Product identification 7 3 3 Scope of delivery 8 4 Electrical connection 8 4 1 Connection to handheld device 8 4 2 Connection to handheld device via M12 cable 9 5 Commissioning 9 5...

Страница 4: ...situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective action This...

Страница 5: ...enance of the measuring system may be carried out only by specially trained technical personnel The technical personnel must be authorized by the plant operator to carry out the specified activities T...

Страница 6: ...place safety As the user you are responsible for complying with the following safety conditions Installation specifications Local standards and regulations 2 4 Operational safety Before commissioning...

Страница 7: ...y that it is protected against impact and moisture The original packaging offers the best protection Make sure to comply with the permitted ambient conditions If you have any questions please contact...

Страница 8: ...in this window as well as the product documentation 3 2 3 Manufacturer address Endress Hauser Conducta GmbH Co KG Dieselstra e 24 D 70839 Gerlingen 3 3 Scope of delivery The scope of delivery compris...

Страница 9: ...681 Connect the M12 cable to the handheld device 2 A0041680 Insert the sensor into the Memosens connection of the M12 cable and lock into place 5 Commissioning 5 1 Calibration and adjustment The senso...

Страница 10: ...for this purpose The COY8 oxygen depleting gel creates an oxygen free medium for zero point calibration Prior to sensor zero point calibration check the following Is the sensor signal stable Has the...

Страница 11: ...nance Take all the necessary precautions in time to ensure operational safety 6 1 Maintenance tasks 6 1 1 Cleaning of sensor Dirt on the sensor can impact the measurement and even cause a malfunction...

Страница 12: ...ons for returning devices The product must be returned if repairs or a factory calibration are required or if the wrong product was ordered or delivered To ensure safe professional and swift product r...

Страница 13: ...2 Zero point gel COY8 Zero point gel for oxygen and disinfection sensors Oxygen free and chlorine free gel for the verification zero point calibration and adjustment of oxygen and disinfection measuri...

Страница 14: ...Oxygen gaseous hPa or Vol Temperature C F 9 1 2 measuring range 0 to 200 SAT Measuring ranges apply for 25 C 77 F and 1013 hPa 15 psi The sensor has a measuring range of up to max 1000 hPa The measure...

Страница 15: ...tance NOTICE Halogen containing solvents ketones and toluene Halogen containing solvents dichloromethane chloroform ketones e g acetone pentanone and toluene have a cross sensitive effect and result i...

Страница 16: ...Technical data Memosens COL37E 16 Endress Hauser Spot cap Stainless steel 1 4435 AISI 316L Spot layer Silicone 9 5 3 Temperature sensor Pt1000 Class A according to DIN IEC 60751...

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ...www addresses endress com 71558605 71558605...

Отзывы: