Memosens CFS51
Maintenance
Hauser
35
10
Maintenance
10.1 Maintenance tasks
L
WARNING
UV radiation from this product
Can cause damage to the eyes and skin!
‣
Avoid any exposure of the eyes and skin to the unshielded product.
‣
When the sensor is switched on, avoid looking directly into the sensor window without
appropriate eye protection. The exposure limits according to IEC 62471:2008 are not
exceeded within the first 100 seconds.
‣
Appropriate protective goggles must be worn to protect against UV radiation.
‣
Cover the light source when performing maintenance tasks that do not need UV light.
L
CAUTION
Acid or medium
Risk of injury, damage to clothing and the system!
‣
Switch off cleaning before the sensor is removed from the medium.
‣
Wear protective goggles and safety gloves.
‣
Clean away splashes on clothes and other objects.
‣
You must perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or log.
The maintenance cycle primarily depends on the following:
• The system
• The installation conditions
• The medium in which measurement takes place
10.1.1 Removing sensor from assembly
1
2
3
A0048246
29
Sensor with assembly
1
M5 screw
2
Ring clip
3
Spacer
To clean or calibrate the sensor, move the sensor to the service position:
1. Shut off the inlet for the process medium before the maintenance tasks.
2. Pay attention to the process pressure and temperature of the medium → 40.