3. Electrical Connection
3.1 General information
Note both the polarity and operating voltage.
Warning!
Do not install, wire up or dismantle the instrument when the power supply is switched
on.
3.2 Connecting the Magphant
Wiring and cable specifications
Wire cross-section
: max. 1.5 mm
2
Cable diameter
: 7...12 mm
Cable gland
: PG 16
Shielded cable should always be used.
Fuse 160 mA slow-blow
Current output 4...20 mA
Load R
L
= 750
Ω
max.
Relay output
75 V DC / 0.5 A
60 V AC / 0.4 A
Power supply 20...30 V DC
Terminal block
b
a025
y
0
2
Signal cable
Power supply cable
Ground terminals
for cable shields
Fig. 12:
Electrical connection