28
Oxy
Oxygen Calibration
Measurement
CAL
ZERO CAL
xxx
nA
PRESS CAL blinks
Place sensor in oxygen‑free medium (e.g.,
nitrogen 5.0) and wait until the measured values
have stabilized.
CAL DATA
Calibration will be performed.
Date
ZERO P. xxx nA
Zero calibration data is displayed.
Date
SLOPE xxx nA
Slope calibration data is displayed.
Automatic return to measuring mode.
ZERO CAL Calibration
(Zero calibration with oxygen-free medium)
The calibration method is selected in the configuration menu.
Please note: To abort calibration, you can press
at any time.