Technical data
Liquiline System CA80NO
164
Hauser
Rockwell
Add-on-Profile Level 3, Faceplate for Factory
Talk SE
IO data
Input (T → O)
Device status and diagnostic message with
highest priority
Measured values:
• 16 AI (analog input) + Unit
• 8 DI (discrete input) + Status
Output (O → T)
Actuating values:
• 4 A0 (analog output) + unit
• 8 DO (discrete output) + Status
15.6 Power supply
Supply voltage
• 100 to 120 V AC / 200 to 240 V AC
or 24 V DC (not available for "Outdoor" version)
• 50 or 60 Hz
Fieldbus connection
Supply voltage: not applicable
Power consumption
All versions except "Outdoor" version
130 VA + 660 VA per hose heater, max. 1450 VA (version with cooling system)
24-V version: max. 105 W
"Outdoor" version
680 VA + 660 VA per hose heater, max. 2000 VA (version with cooling system)
Fuse
5 x 20 mm 10 A/250 V fine-wire fuse for hose trace heating system
Cable entries
• 4 x bores for M16, G3/8, NPT3/8", Memosens connection
• 4 x bores for M20, G1/2, NPT1/2"
Hose entries
4 x bores for M32 for sample inflow and outflow
Cable specification
Cable gland
Permitted cable diameter
M16x1.5 mm
4 to 8 mm (0.16 to 0.32")
M12x1.5 mm (for order version M12 socket for Memosens sensors)
2 to 5 mm (0.08 to 0.20")
M20x1.5 mm
6 to 12 mm (0.24 to 0.48")
NPT³/₈"
4 to 8 mm (0.16 to 0.32")
G³/₈
4 to 8 mm (0.16 to 0.32")
5)
The power value indicated on the nameplate refers to the power consumption when commissioning at 5˚C (41 °F) after an operating time of one
minute.
6)
In the case of the "Outdoor" version, 2 bores are occupied; for this reason only 2 Memosens sensors are possible