Technical data
Liquiline System CA80NO
162
Hauser
Cable specification
Cable type
Recommended: shielded cable
Cable specification
Max. 2.5 mm
2
(14 AWG)
15.4 Relay outputs
Electrical specification
Relay types
• 1 single-pin changeover contact (alarm relay)
• 2 or 4 single-pin changeover contacts (optional with extension modules)
Maximum load
• Alarm relay: 0.5 A
• All other relays: 2.0 A
Relay switching capacity
Base module (Alarm relay)
Switching voltage
Load (max.)
Switching cycles (min.)
230 V AC, cosΦ = 0.8 to 1
0.1 A
700,000
0.5 A
450,000
115 V AC, cosΦ = 0.8 to 1
0.1 A
1,000,000
0.5 A
650,000
24 V DC, L/R = 0 to 1 ms
0.1 A
500,000
0.5 A
350,000
Extension module
Switching voltage
Load (max.)
Switching cycles (min.)
230 V AC, cosΦ = 0.8 to 1
0.1 A
700,000
0.5 A
450,000
2 A
120,000
115 V AC, cosΦ = 0.8 to 1
0.1 A
1,000,000
0.5 A
650,000
2 A
170,000
24 V DC, L/R = 0 to 1 ms
0.1 A
500,000
0.5 A
350,000
2 A
150,000
Minimum load (typical)
• Min. 100 mA at 5 V DC
• Min. 1 mA at 24 V DC
• Min. 5 mA at 24 V AC
• Min. 1 mA at 230 V AC