Operation
Liquiline System CA80HA
142
Hauser
Menu/Setup/Additional functions/Mathematical functions/MF1 to 8/Mode = Redundancy
Function
Options
Info
Deviation control
Selection
• Off
• On
Factory setting
Off
You can monitor the redundancy. Specify an
absolute limit value that must not be exceeded.
Deviation limit
Depends on the selected
measured value
Redundancy
Read only
View this value in a user-defined measuring
screen or output the value via the current
output.
Mathematical function
assignment view
Overview of the configured functions
rH value
To calculate the rH value, a pH sensor and an ORP sensor must be connected. It is irrelevant
whether you are using a pH glass sensor, an ISFET sensor or the pH electrode of an ISE sensor.
Instead of the mathematical functions, you can also connect a combined pH/ORP sensor.
‣
Set the main measured value simply to rH.
Menu/Setup/Additional functions/Mathematical functions/MF1 to 8/Mode = rH calculation
Function
Options
Info
Calculation
Selection
• Off
• On
Factory setting
Off
On/off switch for the function
pH source
Connected pH sensor
Set the input for the pH sensor and the input
for the ORP sensor. Measured value
interrogation is obsolete as you must select pH
or ORP mV.
ORP source
Connected ORP sensor
Calculated rH
Read only
View this value in a user-defined measuring
screen or output the value via the current
output.
Mathematical function
assignment view
Overview of the configured functions
Degassed conductivity
Carbon dioxide from the air can be a contributing factor to the conductivity of a medium. The
degassed conductivity is the conductivity of the medium excluding the conductivity caused by
carbon dioxide.