Diagnostics and troubleshooting
Liquiline System CA80AM
156
Hauser
11.10 System information
Diagnostics/System information
Function
Options
Info
Device tag
Read only
Factory setting
Analyzer_serial number
Individual device tag, --> "General settings"
Order code
Read only
You can order identical hardware with this code.
This code changes on account of changes to the
hardware and you can enter the new code you received
from the manufacturer here
1)
.
To find out what device version you have, enterthe order code in the search screen at the following address:
www.endress.com/order-ident
Orig. order code ext.
Read only
Complete order code for the original device, resulting
from the product structure.
Current order code ext.
Customized text
Current code, taking into account changes to the
hardware. You must enter this code yourself.
Serial number
Read only
The serial number allows you to access device data and
documentation on the Internet:
www.endress.com/device-viewer
Software version
Read only
Current version
Measuring parameter
Read only
Set measuring parameter
MPL version
Read only
Current version
FXAB1 control module
Read only
• Firmware version
• Hardware version
Photometer
Read only
• Firmware version
• Hardware version
• Raw value
• Temperature
Sample preparation 1
Read only
• Operating mode
Depends on the sample preparation
In the case of a two-channel device, Sample preparation
2 is also displayed.
General information
Read only
• Current status
• SP type
• Order code
• Serial number
• Hardware version
• Software version
• Orig. order code ext.
This information is provided for every electronics
module available. Specify the serial numbers and order
codes when servicing, for example.