Degassed conductivity
Carbon dioxide from the air can be a contributing factor to the conductivity of a medium. The
degassed conductivity is the conductivity of the medium excluding the conductivity caused by
carbon dioxide.
In the power station industry, for example, it is advantageous to use the degassed conductivity:
• The percentage of conductivity caused by corrosion products or fouling in the feed water can
be determined as early as when the turbines are started. The system excludes the initially high
conductivity values resulting from the ingress of air.
• If carbon dioxide is regarded as non-corrosive, the live steam can be directed to the turbine far
earlier during startup.
• If the conductivity value increases during normal operation, it is possible to immediately
determine an ingress of coolant or air by calculating the degassed conductivity.
Path: Menu/Setup/Additional functions/Mathematical functions/MF 1 to 6/Mode/Degassed conductivity
Function
Options
Info
Calculation
Options
• Off
• On
Factory setting
Off
On/off switch for the function
Cation conductivity
Connected conductivity
sensor
"Cation conductivity" is the sensor downstream from the
cation exchanger and upstream from the "degassing
module", "Degassed conductivity" is the sensor at the
outlet of the degassing module.
Measured value interrogation is obsolete as you can only
choose conductivity.
Degassed conductivity
Connected conductivity
sensor
CO2 concentration
Read only
View this value in a user-defined measuring screen or
output the value via the current output.