Liquiline System CA80SI
Operation
Hauser
73
Menu/Setup/Analyzer/Calibration
Function
Options
Info
Settings
Nominal concentration
Factory setting
For configuring the concentration of the
calibration standard solution.
Depends on the settings under
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Measuring value
/
Main value
Automatic cleaning
Selection
• Off
• On
Factory setting
For specifying whether cleaning will take place
before every calibration/adjustment (only in the
automatic mode).
10.4 Current inputs
The input can be used as the data source for limit switches and logbooks, for example.
Furthermore, external values can be made available as set points for controllers.
Menu/Setup/Inputs/Current input x:y
Function
Options
Info
Mode
Selection
• Off
• 0 - 20mA
• 4 - 20mA
Factory setting
4 - 20mA
‣
Select the same current range as in the data
source (connected device).
Input variable
Selection
• Parameter
• Current
Factory setting
Current
‣
Select the input variable.
Meas. value format
Selection
• #
• #.#
• #.##
• #.###
Factory setting
#.#
‣
Specify the number of decimal places.
Parameter name
Input variable
=
Parameter
Customized text, 16
characters
‣
Assign a useful name, e.g. the parameter
name that the data source also uses.
Unit of measure
Input variable
=
Parameter
Customized text, 16
characters
You cannot choose the unit from a list. If you
want to use a unit, you must enter it here as
customized text.