2. Calibration for internals with a partially filled silo,

so that a calibration can also be carried out if the probe is partially covered and 
has been shortened.

1.) Go to field 

V9H5

•   Enter 

"333"

 while keeping the 

O

 

key pressed down, (reset to default values)

•   

I

 + 

G

 simultaneously press

2.) Go to field 

V3H0

 (3x 

I

 press)

•   Enter 

"2"

 (partial calibration if the probe is free)

•   Press 

G

 key

3.) Enter 

"1"

  in field 

V3H1

4.) Go to field 

V3H5

 (4x 

G

 press)

•   Enter free length of probe in m, distance to material min. 0.5 m

•   

I

 + 

G

 press simultaneously, the Setup is now evaluated

•   Wait until the bar graph no longer flashes, 

approx. 30 - 60 s

5.) Go to field 

V3H5

•   Enter the entire length of the probe – 0.35 m as lower range-value

6.) Go to field 

V0H1

•   Enter the same value as in 

V3H5

 (zero adjustement)

•   Press 

G

 key

7.) Enter measuring range (span) in field 

V0H2

 , (min. distance to process connection: 300 mm, check value in field 

V3H8

)

3. Operating matrix

The Levelflex is now calibrated.

Checking measurement reliability:

The strength of the signal can be checked after the silo has been filled. Go to field 

V3H2

. If a value >= 3 is shown here then measurement is assured over 

a long period of time. With very shortened probes, the value for the blocking distance should be checked in field 

V3H8

 and compared with the measuring 

range to ensure that the blocking distance lies outside the measuring range.

Additional information: "http://www.endress.com" 

H0

H1

H2

H3

H4

H5

H6

H7

H8

H9

V0
Basic
calibration

Measured value

Technical

unit

[%] (26)

Empty

calibration

[V3H5]

(26)

Full calibration

[0,9 x E]

(26)

Output damping

Seconds

0-250

[5]

(28)

Value for 4 mA

(8mA)

Technical 

unit

[0]

(28)

Value for 20 mA

(16 mA)

Technical 

unit

[100]

(28)

Safety Alarm

0:MIN

1:MAX

2:HOLD

[2]

(28)

Measured distance

(D) meter/feet

(6)

Level

(L1)

meter/feet

(6)

V1

V2
Linaris-
ation

Linearisation

0:Height

1:Activate

2:Table

3:Semi-automatic

4:Clear

5:Linear

[5]

(27)

Line No.

1-11

(27)

Enter level

meter/feet

(27)

Enter

volume

Technical

unit

(27)

Max.-volume

Technical 

unit

(26)

V3
Extended
calibration

Calibration 

Mode

0: Factory calibra-

tion

1: Customer

2: Partial Calibration

[0]

(24)

Calibration

0: Ready

1: Activate

[0]

(24)

Reflection

quality

0...10

(24)

Actual level

m/ft

[E=m]

(-)

Measuring lenght

(range)

E

1...10 m
1...20 m

(24)

Probe type

0: uncoated

1: coated

(-)

Process connection

0: standard

(-)

Blocking distance

Meter/Feet

(35)

Persistence filter

[5]

(34)

V4...V6

not used

V7
Service

*

*

*

*

*

*

*

*

*

V8 
Operating
Mode

Assign current 

output mode 

0: 4...20 mA 1: 4 mA 
limit 2: 4/20 mA 3: 8/

16 mA

[0]

(28)

Select distance

units:

0: m

1: ft

[0]

(26)

Delay time

s

for E 641

[100]

(28)

V9
Simulation

Diagnostic

code

(32)

Last

diagnostic code

(32)

Device and

software

version

(30)

Device

address

(-)

Reset

333: customer

(35)

Simulation

0: off

1: Level

2: Volume
3: current

[0]

(34)

Simulation

value

(34)

Current output

mA

(34)

Security lock

333: unlock

xxx: lock

(22)

VA
Remote
operation

Tag No.

Units after

linearisation

1...12:%, l, hl

m

, dm, cm

qft (=ft

3

), kg, t

ft, US-gal

Material

V99

System