Endress+Hauser iTEMP TMT82 Скачать руководство пользователя страница 13

iTEMP TMT82

Mounting

Hauser

13

120 mm
(4.72 in)

140 mm
(5.51 in)

21 mm
(0.83 in)

!6.5 mm
(0.25 in)

  A0024604

 6

Dimensions of angle bracket for wall mount (complete wall mounting set available as accessory)

Procedure for mounting in a field housing, pos. B:

1. Open the cover (1) of the field housing (4).
2. Guide the mounting screws (2) through the lateral bores in the head transmitter (3).
3. Screw the head transmitter to the field housing.
4. After wiring, close the field housing cover (1) again. →   16

Procedure for mounting on a DIN rail, pos. C:

1. Press the DIN rail clip (4) onto the DIN rail (5) until it engages with a click.
2. Fit the mounting springs on the mounting screws (1) and guide the screws through the

side boreholes of the head transmitter (2). Then fix both mounting screws with the snap

rings (3).

3. Screw the head transmitter (2) onto the DIN rail clip (4).

Содержание iTEMP TMT82

Страница 1: ...ctions they do not replace the Operating Instructions included in the scope of supply Detailed information can be found in the Operating Instructions and the additional documentation Available for all...

Страница 2: ...iTEMP TMT82 2 Endress Hauser Order code Ext ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555...

Страница 3: ...ng requirements 10 4 2 Mounting 11 4 3 Post mounting check 15 5 Electrical connection 16 5 1 Connection conditions 16 5 2 Quick wiring guide 16 5 3 Connecting the sensor cables 17 5 4 Connecting the t...

Страница 4: ...ctional safety Please refer to Safety Manual SD01172T 09 for the use of approved devices in protective systems according to IEC 61508 1 2 Symbols 1 2 1 Safety symbols Symbol Meaning DANGER A0011189 EN...

Страница 5: ...Trained qualified specialists must have a relevant qualification for this specific function and task Are authorized by the plant owner operator Are familiar with federal national regulations Before be...

Страница 6: ...device is used in the hazardous area e g explosion protection or safety equipment Based on the technical data on the nameplate check whether the ordered device is permitted for the intended use in the...

Страница 7: ...te specifications Extended order code with breakdown of the device features on the delivery note Enter the serial number from the nameplate in the W M Device Viewer www endress com deviceviewer All da...

Страница 8: ...symbols 7 Order code and manufacturer ID 6 0044 iTEMP TMT82 Made in Germany 2012 D 87484 Nesselwang TAGXXXXXXXXXXX TAGXXXXXXXXXXX Order code TMT82 xxx xx Ext ord cd XXXXXXXXXXX Ser no XXXXXXXXX Input...

Страница 9: ...requirements of the standards EN 61 010 1 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use and with the EMC requirements as per the IEC EN 61326 series 3 4 1 CE...

Страница 10: ...ls Dimensions in mm in A Spring travel L 5 mm not for US M4 securing screws B Mounting elements for attachable measured value display C Interface for contacting measured value display The same dimensi...

Страница 11: ...tion option JN 1 SIL Mode 40 to 70 C 40 to 158 F Head transmitter in accordance with climate class C1 DIN rail transmitter in accordance with B2 as per EN 60654 1 Condensation as per IEC 60068 2 33 pe...

Страница 12: ...l head cover 8 on the terminal head 2 Guide the connection wires 4 of the insert 3 through the center hole in the head transmitter 5 3 Fit the mounting springs 6 on the mounting screws 7 4 Guide the m...

Страница 13: ...ws 2 through the lateral bores in the head transmitter 3 3 Screw the head transmitter to the field housing 4 After wiring close the field housing cover 1 again 16 Procedure for mounting on a DIN rail...

Страница 14: ...ting screws 6 through the lateral bores of the head transmitter 5 5 Position the head transmitter 5 in the terminal head 4 in such a way that the bus cable terminals 1 and 2 point to the cable entry 6...

Страница 15: ...top A0017821 8 Mounting the DIN rail transmitter 1 Slide the upper DIN rail clip upwards and the lower clip downwards until they click into place 2 Fit the device on the DIN rail from the front 3 Slid...

Страница 16: ...ead transmitter 1 Open the cable gland and the housing cover on the terminal head or the field housing 2 Feed the cables through the opening in the cable gland 3 Connect the cables as shown in 16 If t...

Страница 17: ...ult in the destruction or malfunction of parts of the electronics 5 3 Connecting the sensor cables NOTICE When connecting 2 sensors ensure that there is no galvanic connection between the sensors e g...

Страница 18: ...Fig A solid wire 1 Strip wire end Min stripping length 10 mm 0 39 in 2 Insert the wire end into the terminal 3 Pull the wire gently to ensure it is connected correctly Repeat from step 1 if necessary...

Страница 19: ...er Cable specification A normal device cable suffices if only the analog signal is used A shielded cable is recommended for HART communication Observe grounding concept of the plant In the case of the...

Страница 20: ...d power supply 5 Internal ground connection 6 External ground connection 7 Shielded signal cable recommended for HART protocol The terminals for signal cable connection 1 and 2 are protected against r...

Страница 21: ...connection Notes Does the supply voltage match the information on the nameplate Head transmitter U 11 to 42 VDC DIN rail transmitter U 12 to 42 VDC SIL mode U 11 to 32 VDC for the head transmitter or...

Страница 22: ...splay to the transmitter 6 1 2 Display elements Head transmitter 1 2 3 4 5 6 7 A0008549 15 Optional LC display for head transmitter Item no Function Description 1 Displays the TAG TAG 32 characters lo...

Страница 23: ...message DIN rail transmitter The DIN rail transmitter version does not have an interface to the LC display and therefore does not have a local display either Two LEDs on the front indicate the device...

Страница 24: ...r in the correct position The head transmitter accepts the settings within one second 5 Secure the cover back onto the terminal head or field housing Switching write protection on off Write protection...

Страница 25: ...y with this instruction can result in damage to parts of the electronics The configuration of device specific parameters is described in detail in the Operating Instructions for the device 7 Commissio...

Страница 26: ...configuration If the device is locked and the parameter settings cannot be changed it must first be enabled via the hardware or software lock The device is write protected if the lock symbol appears...

Страница 27: ......

Страница 28: ...www addresses endress com...

Отзывы: