Commissioning
Hauser
11
4.2
Post-connection check
5
Commissioning
5.1
Function check
CAUTION
!
Avoid damage due to incorrect power supply and inaccurate or insecure hose connections
•
Check that all connections are secure. In particular, ensure that all hose connections are secure, so
that no leaks occur.
•
Ensure that the power supply voltage corresponds to the voltage specified on the nameplate.
5.2
Switching on the unit
CAUTION
!
Touching hot surface
Risk of burning because of hot surface!
•
Do not touch the open housing without safety gloves.
NOTICE
Damage to property because of not designated use
•
The device is designated for interval operation, i.e. maximum 3 minutes cleaning, break of at least
six times the cleaning time.
•
The device may not start up against a pressure load and may not be subject to pressure when at a
standstill.
•
Use the unit to transport air only. The cleaning unit should never be used to transport liquids or
aggressive and/or toxic media.
•
Protect the device against splash water and excessive dust.
You can find further information to the programming of the cleaning unit at CM44x in BA450C
(Operation & settings).
Checks
Notes
Is the supply voltage correct?
230 V AC
Are the installed cables strain-relieved and not twisted?
Are all the cable entries installed, tightened and sealed?