that the accuracy of the in-situ adjustment can suffer when just one flow point is defined if
the measured flow is not close to the adjustment value.
Bidirectional flow
Measuring devices that are equipped with the bidirectional option can be adjusted in situ
in both flow directions or in just one flow direction, as required. If the measuring device is
only adjusted in one direction, it is important that the adjustment is in the positive
direction (forward flow) as these adjustment points are automatically replicated to the
negative direction (reverse flow).
Unknown gas composition
If the gas or gas mixture is unknown, or if the gas composition cannot be defined by the
standard gas selection, the user can define the process gas as "Air". This method has the
disadvantage that compensation in the event of variations in the pressure and
temperature cannot be guaranteed. If the operator is not sure of the exact composition of
the gas but can make an approximate guess, it is recommended to use this approximate
gas composition instead of air.
"In-situ adjustment" submenu
Navigation
"Expert" menu → Sensor → In-situ adjustment
‣
In-situ adjustment
Activate in-situ adjustment (17360)
Input type reference value (17351)
Delete values (17355)
Confirm (17356)
Select flow reference (17354)
Stability check (17366)
Actual flow value (17365)
External reference value (17352)
Reference value (17353)
Apply value (17364)
Status (17367)
Description 1 (17359)
Description 2 (17358)