Troubleshooting
Proline Promass 83
88
Hauser
9.2
System error messages
Serious system errors are
always
recognized by the instrument as "Fault message", and are
shown as a lightning flash (
$
) on the display!
"
Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer
for repair. Important procedures must be carried out before you return a flowmeter to
Hauser
Always enclose a duly completed "Declaration of contamination" form. You will find a
preprinted blank of this form at the back of this manual.
!
Note!
See also the information on
No.
Error message / Type
Status signals
(from the
factory,
only HART 7)
Cause
Remedy (spare part
S = System error
$
= Fault message (with an effect on the outputs)
! = Notice message (without any effect on the outputs)
No. # 0xx
→
Hardware error
001
S: CRITICAL FAILURE
$
: # 001
F
Serious device error
Replace the amplifier board.
011
S: AMP HW EEPROM
$: # 011
F
Amplifier:
Defective EEPROM
Replace the amplifier board.
012
S: AMP SW EEPROM
$
: # 012
F
Measuring amplifier: Error when
accessing data of the EEPROM
The EEPROM data blocks in which an error has
occurred are displayed in the TROUBLESHOOTING
function.
Press Enter to acknowledge the errors in question;
default values are automatically inserted instead of
the errored parameter values.
!
Note!
The measuring device has to be restarted if an error
has occurred in a totalizer block (see also error No.
111 / CHECKSUM TOTAL.).
031
S: SENSOR HW DAT
$
: # 031
F
1. S-DAT is not plugged into the
amplifier board correctly (or is
missing).
2. S-DAT is defective.
1. Check whether the S-DAT is correctly plugged into
the amplifier board.
2. Replace the S-DAT if it is defective.
Check that the new, replacement DAT is
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3. Replace measuring electronics boards if necessary.
4. Plug the S-DAT into the amplifier board.
032
S: SENSOR SW DAT
$
: # 032
F
041
S: TRANSM. HW DAT
$
: # 041
F
1. T-DAT is not plugged into the
amplifier board correctly (or is
missing).
2. T-DAT is defective.
1. Check whether the T-DAT is correctly plugged
into the amplifier board.
2. Replace the T-DAT if it is defective.
Check that the new, replacement DAT is
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3. Replace measuring electronics boards if necessary.
4. Plug the T-DAT into the amplifier board.
042
S: TRANSM. SW DAT
$
: # 042
F