Operation
Proline Promass 80
42
Hauser
5.4.5
Device status / Error messages
You can read the extended device status, in this case, current error messages, via Command "48".
The command delivers information which are partly coded in bits (see table below).
!
Note!
You can find a detailed explanation of the device status and error messages and their elimination in
the "System error messages" section
Byte–bit
Error No.
Short error description
0-0
001
Serious device error
0-1
011
Measuring amplifier has faulty EEPROM
0-2
012
Error when accessing data of the measuring amplifier EEPROM
1-1
031
S-DAT: defective or missing
1-2
032
S-DAT: Error accessing saved values
1-5
051
I/O board and the amplifier board are not compatible.
3-3
111
Totalizer checksum error
3-4
121
I/O board and the amplifier board (software versions) are not compatible.
4-3
251
Internal communication fault on the amplifier board.
4-4
261
No data reception between amplifier and I/O board
7-3
351
Current output:
Flow is out of range.
7-4
352
7-5
353
7-6
354
7-7
355
Frequency output:
Flow is out of range.
8-0
356
8-1
357
8-2
358
8-3
359
Pulse output:
Pulse output frequency is out of range.
8-4
360
8-5
361
8-6
362
9-0
379
The measuring tube oscillation frequency is outside the permitted range.
9-1
380
9-2
381
The temperature sensor on the measuring tube is likely defective.
9-3
382
9-4
383
The temperature sensor on the carrier tube is likely defective.
9-5
384
9-6
385
One of the measuring tube sensor coils (inlet or outlet) is likely defective.
9-7
386
10-0
387
One of the measuring tube sensor coils (inlet or outlet) is likely defective.
10-1
388
10-2
389
Amplifier error
10-3
390
12-1
474
Maximum flow value entered is exceeded.
12-7
501
New amplifier software version is loaded. Currently no other commands are
possible.
13-0
502
Upload and download of device files. Currently no other commands are possible.