Gammapilot M FMG60 FOUNDATION Fieldbus
Maintenance and repair
Hauser
81
8
Maintenance and repair
8.1
Cleaning
When cleaning the exterior, always use cleaning agents that do not corrode the surface of
the housing and the seals.
8.2
Repairs
Under the Hauser repair concept, repairs to the Gammapilot M are performed by
Hauser Service in Hauser repair workshops. For more information
contact the Service Department at Hauser.
8.3
Repairs to devices with an Ex-certificate
When repairing devices with an Ex-certificate, please also note the following:
• Repairs to Ex-certified devices may only be carried out by Hauser Service in
Hauser workshops.
• Comply with the prevailing standards, national Ex-area regulations, Safety Instructions
(XA) and certificates.
• Only genuine Hauser spare parts may be used.
• A certified device may only be converted into a different certified device version by
EHauser Service in Hauser workshops.
• All repairs and modifications must be documented.
8.4
Replacement
L
CAUTION
Data upload/download is not permitted if the device is used for safety-related
applications.
‣
After an entire device or an electronics module has been replaced, the parameters can
be downloaded to the device again via the communication interface. For this, the data
must have been uploaded to the PC beforehand using the "FieldCare" software.
Level measurement and point level detection
You can continue measuring without performing a new calibration. However, the
calibration values should be checked as soon as possible since the mounting position may
have changed slightly.
Density and concentration measurement
A new calibration must be performed after the replacement.
8.5
Return
8.5.1
Return
The measuring device must be returned if it is in need of repair or a factory calibration, or
if the wrong measuring device has been delivered or ordered. Legal specifications require
Hauser, as an ISO-certified company, to follow certain procedures when handling
products that are in contact with the medium.