8
Quick programming guide
8.1
Level measurement
Start
Chapter 4, Section 4.1
- Constants for EC 37 Z/EC 47 Z or
FEB 17 / FEB 17 P
- level is displayed at V0H0
Calibration
Chapter 4, Section 4.2 or 4.3
- After calibration level/volume is
displayed at V0H0
Set analogue output signal (optional)
Chapter 6
- Enter settings in the units selected
during calibration or linearization
Lock parameter matrix (optional)
Chapter 4, Section 4.7
Reset parameters:
V9H5
Enter sensor constants
- Zero frequency
V3H5
- Sensitivity
V3H6
Select mode
V8H0
1= Level measurement
Level or volume
V3H0
»Empty« calibration
V0H1
»Full« calibration
V0H2
Select output range
V0H3
0 = 0
…
20 mA/0
…
10 V,
1 = 4
…
20 mA/2
…
10V
Set output damping
V0H4
Set 0/4 mA value
V0H5
Set 20 mA value
V0H6
Set output at fault
V0H7
0 = -10% (-2 mA/-1 V)
1 = +110% (+22 mA/11V)
2 = last measurement
Lock parameter matrix
V8H9
Silometer FMX 570
Chapter 8: Quick programming guide
38