Flowmeter Proline 300
Electrical connection
Hauser
15
Signal cable
Current output 4 to 20 mA HART
A shielded cable is recommended. Observe grounding concept of the plant.
Current output 0/4 to 20 mA
Standard installation cable is sufficient
Pulse /frequency /switch output
Standard installation cable is sufficient
Double pulse output
Standard installation cable is sufficient
Relay output
Standard installation cable is sufficient.
Current input 0/4 to 20 mA
Standard installation cable is sufficient
Status input
Standard installation cable is sufficient
Cable diameter
• Cable glands supplied:
M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in)
• Spring-loaded terminals: Suitable for strands and strands with ferrules.
Conductor cross-section 0.2 to 2.5 mm
2
(24 to 12 AWG).
Requirements for connecting cable – remote display and operating module DKX001
Optionally available connecting cable
Standard cable
2 × 2 × 0.34 mm
2
(22 AWG) PVC cable with common shield (2 pairs, pair-stranded)
Flame resistance
According to DIN EN 60332-1-2
Oil resistance
According to DIN EN 60811-2-1
Shield
Tin-plated copper braid, optical cover ≥ 85 %
Capacitance: core/shield
≤ 200 pF/m
L/R
≤ 24 µH/Ω
Available cable length
5 m (15 ft)/10 m (35 ft)/20 m (65 ft)/30 m (100 ft)
Operating temperature
When mounted in a fixed position: –50 to +105 °C (–58 to +221 °F); when cable can
move freely: –25 to +105 °C (–13 to +221 °F)