Installation
Dipfit W CLA111
16
Hauser
3.4.2
Installation of CLS50
a0007453
Fig. 16:
Removal of the sensor holder and mounting of the sensor CLS50
1
Bayonet lock
2
Sensor holder
3
Gasket
4
Sensor CLS50
Mounting of the sensor:
1.
Push the gasket (pos. 3) over the threaded shaft.
2.
Screw the sensor holder (pos. 2) out of the bayonet lock (pos. 1).
3.
Lead the sensor cable from the bottom through the sensor holder and through the assembly
pipe.
4.
Screw the sensor (pos. 4) from the bottom into the sensor holder.
5.
Screw the sensor holder into the bayonet lock.
3.5
Post-installation check
• After installation, check that all connections are firmly in position and leak-tight.
• When mounting the assembly via pendulum frame, check if the assembly is moving freely.