Cerabar S/Deltabar S/Deltapilot S 4...20mA HART

Description of parameters

Hauser

129

ACK. ALARM (500)
Selection

Acknowledge alarm.

Prerequisite:

• ACK. ALARM MODE = on

Options:

• Abort
• Confirm

The cause of the alarm must be eliminated, the message must be acknowledged via the 
ACK. ALARM parameter and, where applicable, the ALARM DISPL. TIME (



Page 129) 

has to have elapsed before the device starts measuring again following an alarm.



See also these Operating Instructions, Section 8.3 "Confirming messages".

Factory setting:

Abort

RESET ALL ALARMS (603)
Selection

Use this parameter to reset all the messages of the LAST DIAG. CODE parameter.

Options:

• Abort
• Confirm

Factory setting:

Abort

ERROR No.
Entry 

For "Error"-type messages, you can decide whether the device should behave as in the 
event of an alarm (A) or as in the event of a warning (W). Enter the corresponding 
message number for this parameter. 



See also SELECT ALARMTYPE.



See also these Operating Instructions, Section 8.1 "Messages" and Section 8.2 

"Response of outputs to errors".

Prerequisite:

• Digital communication

SELECT ALARMTYPE
(595) – Entry
(600) – Selection

For "Error"-type messages, you can decide whether the device should behave as in the 
event of an alarm (A) or as in the event of a warning (W). 



See also ERROR No.



See also these Operating Instructions, Section 8.2 "Response of outputs to errors".

Options

• Alarm (A): output current assumes a defined value.
• Warning (W): device continues measuring

On-site operation:

1.

Enter the corresponding message number for ERROR No. field.

2.

Select "Alarm" or "Warning" option.

Digital communication:

1.

Enter the corresponding message number via the ERROR No. parameter.

2.

Use the SELECT ALARMTYPE parameter to select the "Alarm" or "Warning" option.

ALARM DELAY (336)
Entry 

Enter alarm response time for all "Error" messages.

!

Note! 

There is no alarm if the cause of the error is eliminated within the alarm delay time.

Input range:

0...100 s

Factory setting:

0.0 s

ALARM DISPL. TIME (480)
Entry 

Enter alarm display time for all "Error" messages. Once the cause of the error is rectified, 
the alarm display time starts running.

!

Note! 

The following applies if the setting for ACK. ALARM MODE = on: 
If an alarm appears and the alarm display time elapses before the alarm has been 
acknowledged, the message will be cleared once it has been acknowledged.



See also these Operating Instructions, Section 8.3 "Confirming messages".

Input range:

0...999.9 s

Factory setting:

0.0 s

Table 33: (GROUP SELECTION 

) OPERATING MENU 



DIAGNOSTICS 



MESSAGES

Parameter name

Description

Содержание Deltabar S PMD70

Страница 1: ...1 valid from Software version 02 20 zz Operating Instructions Description of Instrument Functions Cerabar S PMC71 PMP71 75 Deltabar S FMD76 77 78 PMD70 75 Deltapilot S FMB70 Process pressure Differential pressure Flow Hydrostatic ...

Страница 2: ...rabar S Deltabar S Deltapilot S 4 20mA HART 2 Endress Hauser A0023555 TAG No XXX000 Ser No X000X000000 Order code 00X00 XXXX0XX0XXX www endress com deviceviewer Endress Hauser Operations App Serial number ...

Страница 3: ...Calibration without reference pressure 12 5 Level measurement 14 5 1 Overview of level measurement 14 5 2 Level Easy Pressure level selection 15 5 3 Level Easy Height level selection 19 5 4 Level Standard level selection Linear level type 24 5 5 Level Standard level selection Pressure Linearized level type 28 5 6 Level Standard level selection Height Linearized level type 33 6 Flow measurement 40 ...

Страница 4: ...page reference link takes you to the parameter in question In the Operating program additional parameters and to an extent other parameters are displayed These parameters are not listed in Section 2 You can find these parameters by means of the index See also Section 1 3 1 2 Finding function group using graphic representation All the function groups are shown in table form in Section 3 The page re...

Страница 5: ...WNLOAD SELECT 126 015 FULL PRESSURE See 1 1 See Safety Manual SD00189P for Deltabar S SD00190P for Cerabar S and SD00213P for Deltapilot S 016 EMPTY PRESSURE See 1 017 FULL CALIB See 1 018 EMPTY CALIB See 1 020 LEVEL SELECTION 46 021 SET LRV See 1 022 SET URV See 1 023 OUTPUT UNIT Level Selection Level Easy Pressure 60 023 OUTPUT UNIT Level Selection Level Easy Height 64 025 PROCESS DENSITY Level ...

Страница 6: ...TOTAL 2 109 331 RESET TOTALIZER1 109 332 Pmin ALARM WINDOW 130 333 Pmax ALARM WINDOW 130 334 Tmin ALARM WINDOW 130 335 Tmax ALARM WINDOW 130 336 ALARM DELAY 129 339 DISPLAY CONTRAST 111 342 SET MAX ALARM 113 343 SET MIN CURRENT 113 345 BUS ADDRESS 115 350 DEVICE DESIGN 117 351 DEVICE TYPE Deltabar S 115 352 CONFIG RECORDER 117 354 DEVICE SERIAL No 116 357 PCB TEMPERATURE 117 358 ALLOWED MIN TEMP 1...

Страница 7: ...TABLE LINE NUMB enter values 102 550 EDITOR TABLE X VAL enter values 102 551 EDITOR TABLE Y VAL enter values 102 102 563 POS INPUT VALUE 48 50 or 54 564 LAST DIAG CODE 128 570 Pmax PROC CONN 117 571 MASS FLOW UNIT 94 581 SENSOR MEAS TYPE 120 584 SENSOR PRESSURE Pressure measuring mode 122 SENSOR PRESSURE Level measuring mode 122 SENSOR PRESSURE Flow measuring mode 123 585 HART VERSION 114 591 MINI...

Страница 8: ...level type 70 or 76 CUSTOMER UNIT H Height Linearized level type 84 or 89 708 HEIGHT UNIT Linear level type 70 or 76 HEIGHT UNIT Height Linearized level type 83 or 88 709 MASS UNIT Linear level type 72 MASS UNIT Pressure Linearized level type 80 MASS UNIT Height Linearized level type 85 710 EMPTY PRESSURE Linear level type 74 EMPTY PRESSURE Height Linearized level type 87 711 FULL PRESSURE Linear ...

Страница 9: ...KSTATE See 1 1 See Safety Manual SD00189P for Deltabar S SD00190P for Cerabar S and SD00213P for Deltapilot S SAFETY LOCK 838 SAFETY PASSWORD See 1 840 DIGIT SETS 111 841 DIGIT SETS See 1 844 ACK ALARM MODE See 1 845 MEASURING MODE See 1 847 CALIB OFFSET See 1 848 MAX FLOW See 1 849 MAX PRESS FLOW See 1 850 LOW FLOW CUT OFF See 1 851 SET L FL CUT OFF See 1 852 SET LRV See 1 853 SET URV See 1 854 L...

Страница 10: ...UP pressure 55 BASIC SETUP level Level Easy Pressure 58 BASIC SETUP level Level Easy Height 62 BASIC SETUP level Level Easy Standard 67 BASIC SETUP flow 91 EXTENDED SETUP pressure 96 EXTENDED SETUP level 96 EXTENDED SETUP flow 98 LINEARISATION on site display 100 LINEARISATION Digital communication 104 TOTALIZER SETUP 107 SAFETY CONFIRM See 1 DISPLAY 558 110 OUTPUT 559 111 TRANSMITTER INFO 560 HAR...

Страница 11: ...eady installed for example Note See also Operating Instructions Deltabar S BA00270P Section 6 6 Differential pressure measurement Cerabar S BA00271P Section 6 4 Pressure measurement or Deltapilot S BA00332P Section 6 5 Pressure measurement For a description of the parameters mentioned see Page 45 Table 2 MEASURING MODE Page 53 Table 6 POSITION ADJUSTMENT Page 55 Table 7 BASIC SETUP For a descripti...

Страница 12: ... step 6 2 See table step 7 2 Carry out position adjustment if necessary See Page 53 Table 6 POSITION ADJUSTMENT 3 If necessary select the Pressure measuring mode via the MEASURING MODE parameter On site display Menu path GROUP SELECTION MEASURING MODE Digital communication Menu path OPERATING MENU SETTINGS BASIC SETUP MEASURING MODE 4 On site display Select BASIC SETUP function group Menu path GRO...

Страница 13: ...so perform calibration without reference pressure by means of the QUICK SETUP menu See Page 47 ff Table 3 QUICK SETUP menu You can also specify a customer specific unit See parameter description for PRESS ENG UNIT Page 56 Description 1 If necessary select the Pressure measuring mode via the MEASURING MODE parameter On site display Menu path GROUP SELECTION MEASURING MODE Digital communication Menu...

Страница 14: ...nd mass units are possible The measured value display and the LEVEL BEFORE LIN parameter show the measured value The measured variable is not in direct proportion to the measured pressure as for example with containers with a conical outlet A linearisation table must be entered for the calibration LEVEL SELECTION Level standard LEVEL MODE Pressure Linearized Via LINd MEASURAND parameter Pressure P...

Страница 15: ...MODE Page 53 Table 6 POSITION ADJUSTMENT Page 58 Table 8 LEVEL SELECTION Level Easy Pressure For a description of further relevant parameters see Page 96 Table 16 EXTENDED SETUP Page 122 Table 28 PROCESS VALUES Warning If the measuring mode is changed the span setting URV must be verified in the Calibration Basic Setup operating menu and if necessary reconfigured Description 1 Deltabar S Before yo...

Страница 16: ... for example 7 Select a level unit via the OUTPUT UNIT parameter here m for example 8 Select the Wet option by means of the CALIBRATION MODE parameter 9 Hydrostatic pressure for the lower calibration point is present at the device here 0 mbar for example Select EMPTY CALIB parameter Enter the level value here 0 m for example Confirm the value to assign the pressure value present to the lower level...

Страница 17: ...ectly Due to the orientation of the device there may be a shift in the measured value i e when the container is empty or partly filled the MEASURED VALUE parameter does not display zero To perform a position adjustment see also Page 53 Table 6 Position adjustment For a description of the parameters mentioned see Page 45 Table 2 MEASURING MODE Page 58 Table 8 LEVEL SELECTION Level Easy Pressure For...

Страница 18: ... the Dry option by means of the CALIBRATION MODE parameter 7 Enter the volume value for the lower calibration point via the EMPTY CALIB parameter here 0 l for example 8 Enter the pressure value for the lower calibration point via the EMPTY PRESSURE parameter here 50 mbar 0 75 psi for example 9 Enter the volume value for the upper calibration point via the FULL CALIB parameter here 1000 l 264 gal f...

Страница 19: ...pilot S BA00332P Section 6 4 Level measurement The values entered for EMPTY CALIB FULL CALIB EMPTY PRESSURE FULL PRESSURE EMPTY HEIGHT FULL HEIGHT and SET LRV SET URV must have a minimum interval of 1 for the Level Easy Height level mode The value will be rejected with a warning message if the values are too close together Further limit values are not checked i e the values entered must be appropr...

Страница 20: ...e Page 53 Table 6 POSITION ADJUSTMENT 3 Select the Level measuring mode via the MEASURING MODE parameter On site display Menu path GROUP SELECTION MEASURING MODE Digital communication Menu path OPERATING MENU SETTINGS BASIC SETUP MEASURING MODE 4 If necessary select the Level Easy Height level mode using the LEVEL SELECTION parameter On site display Menu path GROUP SELECTION MEASURING MODE Level L...

Страница 21: ...TY parameter here kg dm3 for example 12 Enter the volume value for the lower calibration point via the EMPTY CALIB parameter here 0 l for example The currently measured hydrostatic pressure is displayed as height here 0 5 m 1 6 ft for example Note To accept the value displayed you must first switch to the Edit mode see the Editing values section and then press the E button to save the value 13 Ent...

Страница 22: ... Due to the orientation of the device there may be a shift in the measured value i e when the container is empty or partly filled the MEASURED VALUE parameter does not display zero To perform a position adjustment see also Page Table 6 Position adjustment For a description of the parameters mentioned see Page 45 Table 2 MEASURING MODE Page 62 Table 9 LEVEL SELECTION Level Easy Height For a descrip...

Страница 23: ...ameter 8 Select a density unit via the DENSITY UNIT parameter here kg dm3 for example 9 Enter the density of the fluid using the ADJUST DENSITY parameter here kg dm3 for example 10 Enter the volume value for the lower calibration point via the EMPTY CALIB parameter here 0 l for example 11 Enter the height value for the lower calibration point via the EMPTY HEIGHT parameter here 0 5 m 1 6 ft for ex...

Страница 24: ...le 28 PROCESS VALUES Warning If the measuring mode is changed the span setting URV must be verified in the Calibration Basic Setup operating menu and if necessary reconfigured Description 1 Deltabar S before configuring the device for your application the pressure piping must be cleaned and the device filled with fluid See Operating Instructions BA00270P Section 6 5 1 P01 PMC71xxx 19 xx xx xx 006 ...

Страница 25: ... 6 Select a pressure unit via the PRESS ENG UNIT parameter here mbar for example 7 Select the Linear option by means of the LEVEL MODE parameter 8 Select the Level option by means of the LIN MEASURAND parameter 9 Select a level unit via the HEIGHT UNIT parameter here m for example 10 Select the Wet option by means of the CALIBRATION MODE parameter 11 The pressure for the lower calibration point is...

Страница 26: ...ters mentioned see Page 45 Table 2 MEASURING MODE Page 67 Table 10 BASIC SETUP Page 70 Table 11 BASIC SETUP Linear level type For a description of further relevant parameters see Page 96 Table 16 EXTENDED SETUP Page 122 Table 26 PROCESS VALUES Warning If the measuring mode is changed the span setting URV must be verified in the Calibration Basic Setup operating menu and if necessary reconfigured D...

Страница 27: ...thin the editing limits 4 Select a pressure unit via the PRESS ENG UNIT parameter here mbar for example P01 xxxx xxxx 19 xx xx xx 012 Fig 14 Current output calibration 5 See table step 13 6 See table step 14 5 Select the Linear option by means of the LEVEL MODE parameter 6 Select the Volume option by means of the LIN MEASURAND parameter 7 Select a volume unit via the UNIT VOLUME parameter here m3 ...

Страница 28: ... level type Page 100 Table 18 LINEARISATION on site operation Page 104 Table 19 LINEARISATION Digital communication For a description of further relevant parameters see Page 96 Table 16 EXTENDED SETUP Page 122 Table 26 PROCESS VALUES Warning If the measuring mode is changed the span setting URV must be verified in the Calibration Basic Setup operating menu and if necessary reconfigured Description...

Страница 29: ...IN parameter Enter the minimum hydrostatic pressure to be expected here 0 mbar for example 11 Select HYDR PRESS MAX Enter the maximum hydrostatic pressure to be expected Carry out linearisation 12 Change the function group Menu path GROUP SELECTION OPERATING MENU SETTINGS LINEARISATION 13 Select TANK CONTENT MIN parameter Specify the minimum tank contents to be expected here 0 m3 for example 14 Se...

Страница 30: ...point LINE NUMB confirm value displayed X VAL the hydrostatic pressure present is displayed On site display Digital communication The X VAL displayed is saved by confirming the Y value See following line Y VAL HART handheld terminal Confirm X VAL displayed Y VAL enter the volume value here 0 m3 for example and confirm the value 19 On site display If you want to enter another point for the linearis...

Страница 31: ...he distance between two points must be maintained Spans for the Pressure linearized option HYDR PRESS MAX HYDR PRESS MIN TANK CONTENT MAX TANK CONTENT MIN Spans for the Height linearized option LEVEL MAX LEVEL MIN TANK CONTENT MAX TANK CONTENT MIN Note See also Operating Instructions for Deltabar S BA00270P or Cerabar S BA00271P Section 6 5 Level measurement or Deltapilot S BA00332P Section 6 4 Le...

Страница 32: ...rent values the desired values must be entered for SET LRV and SET URV 5 On site display Select the Editor table option by means of the TABLE SELECTION parameter P01 xxxxxxxx 05 xx xx xx 015 P01 xxxxxxxx 05 xx xx xx 016 Fig 16 Manual entry of the linearisation table 1 See Section 5 3 1 table step 9 2 See Section 5 3 1 table step 10 3 See table step 3 4 See table step 4 5 See table steps 5 9 6 See ...

Страница 33: ...SETUP Height Linearized level type Page 100 Table 18 LINEARISATION on site operation Page 104 Table 19 LINEARISATION Digital communication For a description of further parameters see Page 96 Table 16 EXTENDED SETUP Page 122 Table 28 PROCESS VALUES Warning If the measuring mode is changed the span setting URV must be verified in the Calibration Basic Setup operating menu and if necessary reconfigur...

Страница 34: ...the UNIT VOLUME parameter here m3 for example 11 Select LEVEL MIN parameter Enter the minimum level to be expected here 0 m for example 12 Select LEVEL MAX parameter Enter the maximum level to be expected here 3 m 9 8 ft for example 13 Select the Wet option via the CALIBRATION MODE parameter calibration mode for the 1st measured variable 14 The pressure for the lower calibration point is present a...

Страница 35: ...to the height of the 1st point LINE NUMB confirm value displayed X VAL the hydrostatic pressure present is measured and converted to the corresponding level and displayed On site display Digital communication The X VAL displayed is saved by confirming the Y value See following line Y VAL HART handheld terminal Confirm X VAL displayed Y VAL enter the volume value here 0 m3 for example and confirm t...

Страница 36: ...RENT parameter Page 113 to specify whether the current output should depict the 1st or 2nd measured variable Depending on the setting of the ASSIGN CURRENT parameter enter the following values for SET LRV and SET URV ASSIGN CURRENT tank content factory setting value volume value or mass value ASSIGNMENT height level value The following applies for the setting ASSIGN CURRENT Tank content If you ent...

Страница 37: ...oned see Page 45 Table 2 MEASURING MODE Page 67 Table 10 BASIC SETUP Page 83 Table 12 BASIC SETUP Height Linearized level type Page 100 Table 18 LINEARISATION on site operation Page 104 Table 19 LINEARISATION Digital communication For a description of further parameters see Page 96 Table 16 EXTENDED SETUP Page 122 Table 28 PROCESS VALUES Warning If the measuring mode is changed the span setting UR...

Страница 38: ...elect TANK CONTENT MIN parameter Specify the minimum tank contents to be expected here 0 m3 for example 8 Select TANK CONTENT MAX parameter Specify the maximum tank contents to be expected here 5 m3 for example 9 On site display Select the Editor table option by means of the TABLE SELECTION parameter 10 Select the Manual option by means of the LIN EDIT MODE parameter 11 Select the New table option...

Страница 39: ...k content If you enter values for TANK CONTENT MIN Page 101 or 104 and TANK CONTENT MAX Page 101 or 105 the SET LRV and SET URV parameters are also changed If you want to assign values other than those for TANK CONTENT MIN and TANK CONTENT MAX to the lower and upper current values the desired values must be entered for SET LRV and SET URV The following applies for the setting ASSIGN CURRENT Height...

Страница 40: ...g Instructions BA00270P Deltabar S Section 6 4 Flow measurement For a description of the parameters mentioned see Page 45 Table 2 MEASURING MODE Page 53 Table 6 POSITION ADJUSTMENT Page 91 ff Table 12 BASIC SETUP Page 98 ff Table 15 EXTENDED SETUP For a description of further parameters see Page 98 Table 15 EXTENDED SETUP Page 123 Table 29 PROCESS VALUES Warning If the measuring mode is changed th...

Страница 41: ...ith fluid See Operating Instructions BA00270P Section 6 4 1 P01 xxxxxxxx 19 xx xx xx 013 Fig 20 Flow measurement calibration 1 See table step 8 2 See table step 9 3 See the following note point 4 4 See the following note point 4 2 Carry out position adjustment if necessary See Page 53 Table 6 POSITION ADJUSTMENT 3 If necessary select the Flow measuring mode via the MEASURING MODE parameter On site...

Страница 42: ...If you enter a value for MAX FLOW the SET URV parameter is also changed If you want to assign values other than zero and MAX FLOW to the lower and upper current values the desired values must be entered for SET LRV and SET URV The following applies for the setting LINEAR SQROOT Differential pres As per the factory settings the lower current value is set to equal zero and the upper current value is...

Страница 43: ...e totalised flow value of the first totalizer The TOTALIZER 2 and TOTAL 2 OVERFLOW parameters display the totalised flow value of the second totalizer See Page 123 ff PROCESS VALUES function group You can use the MENU DESCRIPTOR parameter Page 110 to specify which measured value should be displayed on the on site display Description 1 Calibrate the device as per Section 6 1 2 Change the function g...

Страница 44: ...rameters are appropriately configured For example the EMPTY CALIB parameter is not displayed in the Quick Setup menu Level measuring mode unless the Linear option was selected for the LEVEL MODE parameter and the Wet option was selected for the CALIBRATION MODE parameter There is a comment in the parameter description here stating Note prerequisite LEVEL MODE Linear and CALIBRATION MODE Wet Parame...

Страница 45: ...via digital communication Table 2 GROUP SELECTION MEASURING MODE on site operation Parameter name Description MEASURING MODE 389 Selection Select the measuring mode The operating menu is structured according to the selected measuring mode Note The MEASURING MODE parameter is displayed in the operating program and in the HART handheld terminal in the QUICK SETUP menus and in the BASIC SETUP functio...

Страница 46: ...g the reset code 7864 For additional information see the Safety Manual for Deltabar S SD00189 Cerabar S SD00190 or Deltapilot S SD00213P Options Level Easy Pressure Specify two pressure level value pairs for this level mode The pressure measured value is converted directly to the unit which is selected via the OUTPUT UNIT parameter Page 60 Two calibration modes Wet and Dry are available Wet calibr...

Страница 47: ... SETUP Pressure Parameter name Description This menu displays the most important parameters for the Pressure measuring mode Prerequisite MEASURING MODE Pressure see also Page 45 Note See also Page 55 ff Table 7 BASIC SETUP Page 96 Table 15 EXTENDED SETUP Page 121 ff Table 27 PROCESS VALUES Page 11 ff Section 4 Pressure measurement MEASURING MODE Selection Select the measuring mode The operating me...

Страница 48: ...ample when the container is empty or partly filled the MEASURED VALUE parameter does not display zero or the desired value Example MEASURED VALUE 0 5 mbar 0 0075 psi For the POS INPUT VALUE parameter specify the desired set point for the MEASURED VALUE e g 2 mbar MEASURED VALUEnew POS INPUT VALUE MEASURED VALUE after entry for POS INPUT VALUE 2 0 mbar 0 03 psi The CALIB OFFSET parameter displays t...

Страница 49: ...cription This menu displays the most important parameters for the Level measuring mode Prerequisite MEASURING MODE Level see also Page 45 Note See also Page 67 ff Tables 10 to 13 BASIC SETUP Page 96 ff Table 16 EXTENDED SETUP Page 100 ff Table 18 LINEARISATION on site operation Page 104 ff Table 19 LINEARISATION Digital communication Page 122 ff Table 28 PROCESS VALUES Page 14 ff Section 5 Level m...

Страница 50: ...e desired value Example MEASURED VALUE 0 5 mbar 0 0075 psi For the POS INPUT VALUE parameter specify the desired set point for the MEASURED VALUE e g 2 0 mbar 0 03 psi MEASURED VALUE new POS INPUT VALUE MEASURED VALUE after entry for POS INPUT VALUE 2 0 mbar 0 03 psi The CALIB OFFSET parameter displays the resulting pressure difference offset by which the MEASURED VALUE was corrected CALIB OFFSET ...

Страница 51: ...e present at the device In order for the level value to be saved together with the pressure present at the device the entry field for the level value must first be activated using the or key before confirming with the E key This applies also if the level value is to remain unchanged Factory setting 100 0 DAMPING VALUE 247 Entry Enter damping time time constant The damping affects the speed at whic...

Страница 52: ...he device Due to the orientation of the device there may be a shift in the measured value i e when the container is empty or partly filled the MEASURED VALUE parameter does not display zero Example MEASURED VALUE 2 2 mbar 0 033 psi Correct the MEASURED VALUE via the POS ZERO ADJUST parameter with the Confirm option This means that you are assigning the value 0 0 to the pressure present MEASURED VA...

Страница 53: ...d value and current output react to a change in the pressure Input range 0 0 999 0 s Factory setting 2 0 s or as per order specifications Table 5 GROUP SELECTION QUICK SETUP Flow Parameter name Description 1 1 LANGUAGE BASIC SETUP DISPLAY SETTINGS POSITION ADJUSTMENT OPERATING MENU GROUP SELECTION Measured value EXTENDED SETUP CALIB OFFSET POS ZERO ADJUST POS INPUT VALUE 1 Display via on site disp...

Страница 54: ...ED VALUE 0 5 mbar 0 0075 psi For the POS INPUT VALUE parameter specify the desired set point for the MEASURED VALUE e g 2 0 mbar 0 03 psi MEASURED VALUEnew POS INPUT VALUE MEASURED VALUE after entry for POS INPUT VALUE 2 0 mbar 0 03 psi The CALIB OFFSET parameter displays the resulting pressure difference offset by which the MEASURED VALUE was corrected CALIB OFFSET MEASURED VALUEold POS INPUT VAL...

Страница 55: ...ay via on site display only 2 Display via digital communication Table 7 GROUP SELECTION OPERATING MENU SETTINGS BASIC SETUP Pressure Parameter name Description Prerequisite MEASURING MODE Pressure see also Page 45 Note See also Page 47 Table 3 QUICK SETUP Page 96 Table 15 EXTENDED SETUP Page 121 ff Table 27 PROCESS VALUES Page 11 ff Section 4 Pressure measurement MEASURING MODE Selection Select th...

Страница 56: ...cified as the customer specific unit crate is displayed If the unit contains a slash up to eight characters can be shown on the on site display The maximum number of characters in the counter is again limited to five For example if crates m2 is specified as the customer specific unit crate m2 is displayed In the FieldCare all eight characters are displayed In the HART handheld terminal the custome...

Страница 57: ...ed in the bottom line Options Abort Confirm GET URV 310 Entry Set upper range value reference pressure is present at device The pressure for the upper current value 20 mA is present at device With the Confirm option you assign the upper current value to the pressure value present On site display the pressure value present is displayed in the bottom line Options Abort Confirm DAMPING VALUE 247 Entr...

Страница 58: ...SS ENG UNIT CUSTOMER UNIT P CUST UNIT FACT P BASIC SETUP DISPLAY SETTINGS POSITION ADJUSTMENT OPERATING MENU GROUP SELECTION Measured value MEASURING MODE EXTENDED SETUP 1 Display via on site display only 2 Display via digital communication EMPTY CALIB FULL CALIB LEVEL SELECTION OUTPUT UNIT CALIBRATION MODE Table 8 GROUP SELECTION OPERATING MENU SETTINGS BASIC SETUP Level LEVEL SELECTION Level Eas...

Страница 59: ...first five characters are shown on the on site display For example if crates is specified as the customer specific unit crate is displayed If the unit contains a slash up to eight characters can be shown on the on site display The maximum number of characters in the counter is again limited to five For example if crates m2 is specified as the customer specific unit crate m2 is displayed In the Fie...

Страница 60: ... a value for this parameter you assign a level volume mass or percentage value to the pressure present at the device The unit is selected via the OUTPUT UNIT parameter Page 60 Prerequisite CALIBRATION MODE Wet Note For this parameter the on site display shows the level value to be entered and the pressure present at the device In order for the level value to be saved together with the pressure pre...

Страница 61: ...re level value pair for the upper calibration point The unit is selected via the OUTPUT UNIT parameter Page 60 Prerequisite CALIBRATION MODE Dry Factory setting 100 0 FULL PRESSURE 005 Entry Enter pressure value for the upper calibration point container full See also FULL CALIB Prerequisite CALIBRATION MODE Dry Factory setting 100 0 SET LRV 013 Entry Enter value for the lower current value 4 mA Th...

Страница 62: ... OPERATING MENU GROUP SELECTION Measured value MEASURING MODE EXTENDED SETUP 1 Display via on site display only 2 Display via digital communication LEVEL SELECTION OUTPUT UNIT CALIBRATION MODE HEIGHT UNIT FULL HEIGHT DENSITY UNIT ADJUST DENSITY EMPTY CALIB FULL CALIB FULL CALIB Table 9 GROUP SELECTION OPERATING MENU SETTINGS BASIC SETUP Level LEVEL SELECTION Level Easy Height The following paramet...

Страница 63: ...first five characters are shown on the on site display For example if crates is specified as the customer specific unit crate is displayed If the unit contains a slash up to eight characters can be shown on the on site display The maximum number of characters in the counter is again limited to five For example if crates m2 is specified as the customer specific unit crate m2 is displayed In the Fie...

Страница 64: ...ration takes place by filling and emptying the container The measured pressure is converted to the chosen height unit using the HEIGHT UNIT DENSITY UNIT and ADJUST DENSITY parameters In the case of two different levels the level volume mass or percentage value entered is assigned to the converted height value Dry Dry calibration is a theoretical calibration For this calibration specify two height ...

Страница 65: ...The unit is selected via the OUTPUT UNIT parameter Page 64 Prerequisite CALIBRATION MODE Wet Note For this parameter the on site display shows the level value to be entered and the pressure present at the device In order for the level value to be saved together with the pressure present at the device the entry field for the level value must first be activated using the or key before confirming wit...

Страница 66: ...r percentage value for the upper current value 20 mA The unit is selected via the OUTPUT UNIT parameter Page 64 Factory setting 100 0 DAMPING VALUE 247 Entry Enter damping time time constant The damping affects the speed at which all subsequent elements such as the on site display measured value and current output react to a change in the pressure Input range 0 0 to 999 0 s Factory setting 2 0 s o...

Страница 67: ... the span setting URV must be verified in the Calibration Basic Setup operating menu and if necessary reconfigured Prerequisite Digital communication Options Pressure Level Deltabar S Flow Factory setting Pressure PRESS ENG UNIT 060 Selection Select pressure unit If a new pressure unit is selected all pressure specific parameters are converted and displayed with the new unit Options mbar bar mmH2O...

Страница 68: ...value to be displayed in PU PU packing unit MEASURED VALUE 10000 Pa i 1 PU Entry CUSTOMER UNIT P PU Entry CUST UNIT FACT P 0 0001 Result MEASURED VALUE 1 PU Factory setting 1 0 LEVEL MODE 718 Selection Select level type Options Linear the measured variable level volume mass or is in direct proportion to the measured pressure See also Page 70 ff Table 9 Pressure Linearized the measured variable vol...

Страница 69: ...Height Linearized Flow Pressure Level TANK VOLUME TANK HEIGHT SET LRV SET URV MASS UNIT CUSTOMER UNIT M CUST UNIT FACT M DAMPING VALUE EMPTY CALIB DENSITY UNIT ADJUST DENSITY HEIGHT UNIT 100 POINT ZERO POSITION CUSTOMER UNIT H CUST UNIT FACT H FULL PRESSURE EMPTY PRESSURE FULL CALIB ADJUSTED DENSITY CALIBRATION MODE HEIGHT UNIT CUSTOMER UNIT H CUST UNIT FACT H VOLUME UNIT CUSTOMER UNIT V CUST UNIT...

Страница 70: ... unit Prerequisite LIN MEASURAND Level Options mm cm dm m inch ft User unit see also the following parameter description for CUSTOMER UNIT H and CUST UNIT FACT H Factory setting m CUSTOMER UNIT H 706 Entry Enter text unit for customer specific level unit You can enter a maximum of eight alphanumeric characters here See also CUST UNIT FACT H Prerequisite LIN MEASURAND Level HEIGHT UNIT User unit No...

Страница 71: ... UNIT V 608 Entry Enter text unit for customer specific volume unit You can enter a maximum of eight alphanumeric characters here See also CUST UNIT FACT V Prerequisite LIN MEASURAND Volume UNIT VOLUME User unit Note Only the first five characters are shown on the on site display For example if crates is specified as the customer specific unit crate is displayed If the unit contains a slash up to ...

Страница 72: ...try Enter text unit for customer specific mass unit You can enter a maximum of eight alphanumeric characters here See also CUST UNIT FACT M Prerequisite LIN MEASURAND Mass MASS UNIT User unit Note Only the first five characters are shown on the on site display For example if crates is specified as the customer specific unit crate is displayed If the unit contains a slash up to eight characters can...

Страница 73: ...measured variable the tank volume and the tank height must be entered see Page 75 TANK VOLUME and TANK HEIGHT The density must also be entered in the case of a zero point shift level offset see Page 74 ADJUST DENSITY For the measured variable the density of the fluid must be entered and a level assigned to the 100 point see Page 74 and 77 ADJUST DENSITY and 100 POINT If the measurement should not ...

Страница 74: ...lies also if the level value is to remain unchanged Factory setting 100 0 FULL PRESSURE 711 Display Displays the pressure value for the upper calibration point container full See also FULL CALIB Prerequisite CALIBRATION MODE Wet Factory setting High sensor limit see PRESS SENS HILIM Page 119 ADJUSTED DENSITY 810 Display Displays the density calculated from the upper and lower level point Prerequis...

Страница 75: ...o five For example if crates m2 is specified as the customer specific unit crate m2 is displayed In the FieldCare all eight characters are displayed In the HART handheld terminal the customer specific unit is only displayed in the CUSTOMER UNIT H parameter The measured value is displayed with the additional text User Unit Factory setting _ _ _ _ _ _ _ _ CUST UNIT FACT V 607 Entry Enter conversion ...

Страница 76: ...ve For example if crates m2 is specified as the customer specific unit crate m2 is displayed In the FieldCare all eight characters are displayed In the HART handheld terminal the customer specific unit is only displayed in the CUSTOMER UNIT H parameter The measured value is displayed with the additional text User Unit Factory setting _ _ _ _ _ _ _ _ CUST UNIT FACT H 705 Entry Enter conversion fact...

Страница 77: ...x xx xx 001 Fig 32 Zero point shift 1 Device is mounted above the level lower range value a positive value has to be entered for ZERO POSITION 2 Device is mounted below the level lower range value a negative value has to be entered for ZERO POSITION SET LRV 719 Entry Enter level value for the lower current value 4 mA Factory setting 0 0 SET URV 720 Entry Enter level value for the upper current val...

Страница 78: ...display only 2 Display via digital communication Parameters marked with an asterisk are not always displayed The display depends on the settings of other parameters see the following Table Table 12 GROUP SELECTION OPERATING MENU SETTINGS BASIC SETUP Level LEVEL MODE Pressure Linearized Parameter name Description The following parameters are displayed if you selected the Pressure Linearized option ...

Страница 79: ...the unit contains a slash up to eight characters can be shown on the on site display The maximum number of characters in the counter is again limited to five For example if crates m2 is specified as the customer specific unit crate m2 is displayed In the FieldCare all eight characters are displayed In the HART handheld terminal the customer specific unit is only displayed in the CUSTOMER UNIT V pa...

Страница 80: ... In the FieldCare all eight characters are displayed In the HART handheld terminal the customer specific unit is only displayed in the CUSTOMER UNIT M parameter The measured value is displayed with the additional text User Unit Factory setting _ _ _ _ _ _ _ _ CUST UNIT FACT M 703 Entry Enter conversion factor for a customer specific mass unit The conversion factor must be entered in relation to th...

Страница 81: ...o be expected the more accurate the measurement result Factory setting High sensor limit See PRESS SENS HILIM Page 119 DAMPING VALUE 247 Entry Enter damping time time constant The damping affects the speed at which all subsequent elements such as the on site display measured value and current output react to a change in the pressure Input range 0 0 999 0 s Factory setting 2 0 s or as per order spe...

Страница 82: ...rized Height Linearized Flow Pressure Level MASS UNIT CUSTOMER UNIT V CUST UNIT FACT M DAMPING VALUE EMPTY CALIB DENSITY UNIT ADJUST DENSITY HEIGHT UNIT 100 POINT ZERO POSITION CUSTOMER UNIT H CUST UNIT FACT H FULL PRESSURE EMPTY PRESSURE FULL CALIB ADJUSTED DENSITY CALIBRATION MODE HEIGHT UNIT CUSTOMER UNIT H CUST UNIT FACT H COMB MEASURAND LEVEL MODE PRESS ENG UNIT CUSTOMER UNIT P CUST UNIT FACT...

Страница 83: ...sured variable The 2nd measured variable is assigned to the 1st measured variable by means of this table Prerequisite MEASURING MODE Level see also Page 45 LEVEL SELECTION Level Standard see also Page 46 LEVEL MODE Height Linearized see also Page 68 Note See also Page 67 ff Table 10 BASIC SETUP general Page 96 ff Table 16 EXTENDED SETUP Page 100 ff Table 18 LINEARISATION on site operation Page 104...

Страница 84: ...is only displayed in the CUSTOMER UNIT H parameter The measured value is displayed with the additional text User Unit Factory setting _ _ _ _ _ _ _ _ CUST UNIT FACT H 705 Entry Enter conversion factor for a customer specific level unit The conversion factor must be entered in relation to the SI unit m See also CUSTOMER UNIT H Prerequisite COMB MEASURAND Height and Volume HEIGHT UNIT User unit COMB...

Страница 85: ...omer specific unit is only displayed in the CUSTOMER UNIT V parameter The measured value is displayed with the additional text User Unit Factory setting _ _ _ _ _ _ _ _ CUST UNIT FACT V 607 Entry Enter conversion factor for a customer specific volume unit The conversion factor must be entered in relation to the SI unit m3 See also CUSTOMER UNIT V Prerequisite COMB MEASURAND Height and Volume HEIGH...

Страница 86: ...ket Entry CUSTOMER UNIT M bucket Entry CUST UNIT FACT M 0 1 Result MEASURED VALUE 1 bucket Factory setting 1 0 LEVEL MIN 755 Entry Enter the minimum level to be expected The input limits for the calibration editing limits are derived from the value entered The closer the value entered corresponds to the minimum level to be expected the more accurate the measurement result Note The following applie...

Страница 87: ...ing the calibration mode See also Page 97 Factory setting Wet EMPTY CALIB 314 Entry Enter level value for the lower calibration point container empty The container is either empty or part full By entering a value for this parameter you are assigning a level value to the pressure present at the device See also EMPTY PRESSURE Prerequisite CALIBRATION MODE Wet Factory setting 0 0 EMPTY PRESSURE 710 D...

Страница 88: ...B MEASURAND Height and CALIBRATION MODE Dry COMB MEASURAND Height and Volume CALIBRATION MODE Dry COMB MEASURAND Height and Mass CALIBRATION MODE Dry COMB MEASURAND Height and CALIBRATION MODE Dry COMB MEASURAND Height and Volume CALIBRATION MODE Dry COMB MEASURAND Height and Mass CALIBRATION MODE Dry Factory setting 1 0 HEIGHT UNIT 708 Selection Select level unit Prerequisite COMB MEASURAND Heigh...

Страница 89: ...NIT H parameter The measured value is displayed with the additional text User Unit Factory setting _ _ _ _ _ _ _ _ CUST UNIT FACT H 705 Entry Enter conversion factor for a customer specific level unit The conversion factor must be entered in relation to the SI unit m See also CUSTOMER UNIT H Prerequisite COMB MEASURAND Height and Volume CALIBRATION MODE Dry HEIGHT UNIT User unit COMB MEASURAND Hei...

Страница 90: ...lower range value a positive value has to be entered for ZERO POSITION 2 Device is mounted below the level lower range value a negative value has to be entered for ZERO POSITION Factory setting 0 0 DAMPING VALUE 247 Entry Enter damping time time constant The damping affects the speed at which all subsequent elements such as the on site display measured value and current output react to a change in...

Страница 91: ...ALUE MAX FLOW MAX PRESS FLOW CUST UNIT FACT F CUSTOMER UNIT F LANGUAGE Flow Pressure Level FLOW MEAS TYPE PRESS ENG UNIT CUSTOMER UNIT P CUST UNIT FACT P BASIC SETUP DISPLAY SETTINGS POSITION ADJUSTMENT OPERATING MENU GROUP SELECTION Measured value MEASURING MODE EXTENDED SETUP UNIT FLOW NORM FLOW UNIT STD FLOW UNIT MASS FLOW UNIT Table 14 GROUP SELECTION OPERATING MENU SETTINGS BASIC SETUP Flow P...

Страница 92: ... of 4 C 39 2 F 2 The conversion factor of the pressure units refers to a reference temperature of 0 C 32 F Factory setting Depends on the sensor nominal measuring range mbar or bar or as per order specifications CUSTOMER UNIT P 075 Entry Enter text unit for customer specific pressure unit You can enter a maximum of eight alphanumeric characters here See also CUST UNIT FACT P Prerequisite PRESS ENG...

Страница 93: ...t When a new flow unit is selected all flow specific parameters are converted and displayed with the new unit within a flow mode FLOW MEAS TYPE When the flow mode is changed conversion is not possible Prerequisite FLOW MEAS TYPE Volume p cond Options m3 s m3 min m3 h m3 day l s l min l h hl s hl min hl day ft3 s ft3 min ft3 h ft3 day ACFS ACFM ACFH ACFD ozf s ozf min US Gal s US Gal min US Gal h U...

Страница 94: ...n h ton day User unit see also the following parameter description for CUSTOMER UNIT F and CUST UNIT FACT F Factory setting kg s CUSTOMER UNIT F 610 Entry Enter text unit for customer specific flow unit You can enter a maximum of eight alphanumeric characters here See also CUST UNIT FACT F Prerequisite UNIT FLOW User unit NORM FLOW UNIT User unit STD FLOW UNIT User unit MASS FLOW UNIT User unit No...

Страница 95: ...o assign a flow to the upper current value If you want to assign the upper current value a value other than that for MAX FLOW you must enter the desired value for SET URV SET URV Page 100 Factory setting 1 0 MAX PRESS FLOW 634 Entry Enter maximum pressure of primary element See layout sheet of primary element This value is assigned to the maximum flow value see MAX FLOW Note Use the LINEAR SQROOT ...

Страница 96: ...RV TEMP ENG UNIT 1 Display via on site display only Parameters marked with an asterisk are not always displayed The display depends on the settings of other parameters see the following Table Table 15 GROUP SELECTION OPERATING MENU SETTINGS EXTENDED SETUP Pressure Parameter name Description Prerequisite MEASURING MODE Pressure see also Page 45 Note See also Page 11 ff Section 4 Pressure measuremen...

Страница 97: ... of a residual volume measurement a negative value shall be entered for this parameter Factory setting 1 0 PROCESS DENSITY 025 811 Entry Enter a new density value for density correction The calibration was carried out with the medium water for example Now the container is to be used for another fluid with another density The calibration is corrected appropriately by entering the new density value ...

Страница 98: ... URV 763 Entry Enter value for the upper current value 20 mA Prerequisite LEVEL MODE Pressure Linearized or Height Linearized Note For the LEVEL MODE Height Linearized you can use the ASSIGN CURRENT parameter Page 113 to specify whether the current output should depict the 1st or 2nd measured variable height or tank content Depending on the setting of the ASSIGN CURRENT parameter enter the followi...

Страница 99: ...tting Off SET L FL CUT OFF 323 Entry Enter switch off point of low flow cut off The hysteresis between the switch on point and the switch off point is always 1 of the end flow value See also LOW FLOW CUT OFF Prerequisite LOW FLOW CUT OFF on Input range Switch off point 0 50 of end flow value MAX FLOW P01 PMD7xxxx 05 xx xx xx 000 Factory setting 5 of end flow value SET LRV 637 Entry Depending on th...

Страница 100: ...ial pres If you enter a new value for MAX PRESS FLOW the SET URV value is also changed If you want to assign the upper current value a value other than that for MAX PRESS FLOW you must enter the desired value for SET URV MAX PRESS FLOW Page 95 Factory setting MAX FLOW Table 17 GROUP SELECTION OPERATING MENU SETTINGS EXTENDED SETUP Flow Parameter name Description 1 1 EDITOR TABLE EDITOR TABLE MEASU...

Страница 101: ...ue for TANK CONTENT MAX the value for SET URV is also changed Use SET URV to assign a value volume value or mass value to the upper current value If you want to assign the upper current value a value other than that for TANK CONTENT MAX you must enter the desired value for SET URV SET URV Page 98 For the setting LEVEL MODE Height Linearized and ASSIGN CURRENT Linear the TANK CONTENT MAX parameter ...

Страница 102: ...the COMB MEASURAND parameter enter the volume mass or value X VAL the hydrostatic pressure present is measured Depending on the setting in the COMB MEASURAND parameter the measured pressure is converted to a level unit or a and displayed The value is saved by confirming the Y value Factory setting LINE NUMB 1 X VAL 0 0 Y VAL 0 0 EDITOR TABLE Entry manual edit mode LINE NUMB 549 Y VAL 551 X VAL 550...

Страница 103: ...elete point in this case the 5th point for example Select point 5 via the EDITOR TABLE LINE NUMB parameter Using the TABLE EDITOR 770 parameter select the option Delete point The 5th point is deleted All of the following points are pushed up one number i e following deletion the 6th point becomes Point 5 Factory setting Next point MEASURING TABLE 549 Display A point of the linearisation table save...

Страница 104: ... Height Linearized see also Page 68 Note See also Page 14 ff Section 5 Level measurement TANK CONTENT MIN Entry Enter the minimum tank contents to be expected The input limits for the calibration editing limits are derived from the value entered The closer the value entered corresponds to the minimum tank content to be expected the more accurate the measurement result Note If you enter a new value...

Страница 105: ...entering new values Options View meas table Editor table Factory setting View meas table LIN EDIT MODE Selection Select the entry mode for the linearisation table Prerequisite TABLE SELECTION Editor table Options Manual The container neither has to be filled nor emptied for this entry mode Enter the value pairs for the linearisation table Semiautomatic the container is filled or emptied in stages ...

Страница 106: ...on table Prerequisite TABLE SELECTION Editor table Depending on the setting in the LINd MEASURAND or COMB MEASURAND parameters enter a volume mass or value here See also this table parameter description for LIN EDIT MODE LINE NUMB X VAL Manual entry mode X VAL Semiautomatic entry mode EDITOR TABLE Options Select the function for the editor table Prerequisite TABLE SELECTION Editor table Options Ne...

Страница 107: ...table in the Tables window TANK DESCRIPTION Entry Enter tank description max 32 alphanumeric characters Factory setting _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Table 19 GROUP SELECTION OPERATING MENU SETTINGS LINEARISATION Digital communication Parameter name Description 1 1 Flow measuring mode only NEG FLOW TOT 1 RESET TOTALIZER 1 TOTALIZER 2 UNIT TOT 2 USER UNIT FACT U U ...

Страница 108: ...n be shown on the on site display The maximum number of characters in the counter is again limited to five For example if crates m2 is specified as the customer specific unit crate m2 is displayed In the FieldCare all eight characters are displayed In the HART handheld terminal the customer specific unit is only displayed in the TOT 1 USER UNIT parameter The measured value is displayed with the ad...

Страница 109: ...Enter text unit for customer specific unit for totalizer 2 See also TOT 1 USER UNIT Prerequisite TOTALIZER 2 UNIT User unit Factory setting _ _ _ _ _ _ _ _ FACT U U TOTAL 2 330 Selection Enter conversion factor for a customer specific unit for totalizer 2 See also FACT U U TOTAL 1 Prerequisite TOTALIZER 2 UNIT User unit Factory setting 1 0 NEG FLOW TOT 2 416 Selection Specify way of counting negat...

Страница 110: ...00271P Cerabar S or BA00332P Deltapilot S Section 5 1 On site display Options Auto x x x xx x xxx x xxxx x xxxxx Factory setting Auto ALTERNATE DATA 423 Selection Switch on Alternating display mode In this display mode the on site display alternates between the following measured values depending on the measuring mode selected Pressure Main measured value PV Pressure Temperature and Current Level ...

Страница 111: ...13 4 contrast weaker brighter 13 contrast stronger darker Factory setting 8 DIGITS SETS 840 Display This parameter is used to check the correct display of characters and digits on the user interface If the characters and digits are displayed correctly this parameter displays the string 0123456789 Table 21 GROUP SELECTION OPERATING MENU DISPLAY Parameter name Description 1 1 SET MIN CURRENT ASSIGN ...

Страница 112: ...RING MODE Flow with the setting for LINEAR SQROOT Differential pres 695 MEASURING MODE Flow with the setting LINEAR SQROOT Flow square root 696 MEASURING MODE Level LEVEL MODE Linear or Pressure Linearized and LEVEL MODE Height Linearized with the setting for ASSIGN CURRENT Level 764 MEASURING MODE Level LEVEL MODE Height Linearized with the setting for ASSIGN CURRENT Tank content Factory setting ...

Страница 113: ... mA SET MIN CURRENT 343 Entry Enter lower current limit Some switching units sometimes do not accept currents less than 4 0 mA Options 3 8 mA 4 0 mA Factory setting 3 8 mA ASSIGN CURRENT 760 Selection Specify current signal for the Level measuring mode See also SET LRV Page 98 and SET URV Page 98 Prerequisite MEASURING MODE Level LEVEL MODE Height Linearized Options Height Tank content Factory set...

Страница 114: ...IAL DRAIN VENT MAT DRAIN VENT POS THREAD ADDITIONAL INFO SEAL TYPE SENSOR MEAS TYPE BOLTS MATERIAL DEVICE REVISION ELECTR SERIAL No CUST TAG NUMBER LONG TAG NUMBER PROC CONN TYPE MAT PROC CONN MAT PROC CONN PRESS SENS LOLIM PRESS SENS HILIM MINIMUM SPAN HART DATA BUS ADRESS TRANSMITTER INFO TRANSMITTER DATA DEVICE SERIAL No PROCESS CONNECTION Pmax PROC CONN SENSOR DATEN SENSOR SER No TOTALIZER SET...

Страница 115: ... DEVICE TYPE 002 Display Displays the device type in decimal numerical format here Deltapilot S 26 The extended device type is a composition of the manufacturer number 17 and the device type 26 Prerequisite Pressure transmitter Deltapilot S DEVICE REVISION 699 Display Displays the device revision BURST MODE Selection Switches Burst Mode function on and off Selection On Off Prerequisite Digital com...

Страница 116: ...SS SENSOR PRESSURE SENSOR TEMP PCB TEMPERATURE SUPPRESSED FLOW TOTALIZER 1 TOTALIZER 2 LEVEL BEFORE LIN TANK CONTENT Prerequisite Digital communication SECONDARY VALUE Display second process value See also SECONDARY VAL IS Prerequisite Digital communication THIRD VALUE IS Select third process value See also SECONDARY VAL IS Prerequisite Digital communication THIRD VALUE Display third process value...

Страница 117: ...ATURE 357 Display Displays the measured temperature of the main electronics ALLOWED MIN TEMP 358 Display Displays the lower temperature limit of the main electronics ALLOWED MAX TEMP 359 Display Displays the upper temperature limit of the main electronics DIP STATUS 363 Display Displays the status of DIP switch 1 on the electronic insert You can lock or unlock parameters relevant to the measured v...

Страница 118: ...PVDF ECTFE Factory setting As per order specifications MAT PROC CONN 361 Selection For selecting and displaying the material of the process connection P See also parameter description for MAT PROC CONN Prerequisite Deltabar S differential pressure transmitter SEAL TYPE 362 Selection For selecting and displaying the material of the process seal Options Not used Unknown Special FKM Viton NBR EPDM Ur...

Страница 119: ...al communication DIAPHRAG MAT For selecting and displaying the material of the process isolating diaphragm on the positive side Prerequisite Digital communication DIAPHRAG MAT For selecting and displaying the material of the process isolating diaphragm on the negative side Prerequisite Digital communication NR OF REMOTE SEAL For selecting and displaying the number of diaphragm seals Prerequisite D...

Страница 120: ...NE 365 Display Displays the material of the process isolating diaphragm Factory setting As per version in the order code For Deltabar S see Technical Information TI00382P for Cerabar S see Technical Information TI00383P or for Deltapilot S see Technical Information TI00416P Ordering information section FILLING FLUID 366 Display Displays the filling fluid Tmin SENSOR 368 Display Displays the lower ...

Страница 121: ...PRESSURE CORRECTED PRESS SENSOR PRESSURE SENSOR TEMP MEAS VAL TREND LEVEL BEFORE LIN TANK CONTENT MEASURED VALUE Flow Pressure Level PROCESS VALUE PROCESS INFO PEAK HOLD INDICATOR TOTALIZER SETUP LANGUAGE SETTINGS POSITION ADJUSTMENT OPERATING MENU GROUP SELECTION Measured value PRESSURE CORRECTED PRESS SENSOR PRESSURE SENSOR TEMP MEAS VAL TREND PRESSURE CORRECTED PRESS SENSOR PRESSURE SENSOR TEMP...

Страница 122: ...level type this value corresponds to the LEVEL BEFORE LIN parameter In the Level measuring mode with Height Linearized level type this value corresponds to the TANK CONTENT parameter Prerequisite Digital communication On site operation For on site operation the MEASURED VALUE parameter is displayed on the 1st level PRESSURE 301 Display Displays the measured pressure after sensor recalibration posi...

Страница 123: ...Digital communication On site operation For on site operation the MEASURED VALUE parameter is displayed on the 1st level PRESSURE 301 Display Displays the measured pressure after sensor recalibration position adjustment and damping This value corresponds to the MEASURED VALUE parameter in the Pressure measuring mode P01 xMx7xxxx 05 xx xx xx 011 CORRECTED PRESS 434 Display Displays the measured pre...

Страница 124: ...limit value is lower nominal pressure limit of sensor 10 of upper nominal pressure limit of sensor You can reset this counter by means of the RESET PEAKHOLD parameter MIN MEAS PRESS 469 Display Displays the smallest measured pressure value peak hold indicator You can reset this indicator by means of the RESET PEAKHOLD parameter COUNTER T Tmax 404 Display Displays the number of times the specified ...

Страница 125: ...RATING MENU PROCESSINFO PEAK HOLD INDICATOR Parameter name Description I J I J J J J J J J 1 1 2 none Level Pressure Tank content Flow Current Alarm Warning TOTALIZER SETUP LANGUAGE SETTINGS POSITION ADJUSTMENT OPERATING MENU GROUP SELECTION Measured value SIM LEVEL SIM PRESSURE SIM TANK CONT SIM FLOW VALUE SIM CURRENT SIM ERROR NO OPERATION ENTER RESET CODE OPERATING HOURS INSERT PIN NO DIAGNOSTI...

Страница 126: ...ation Options Lock enter a number between 0 9999 which is 100 Unlock enter the number 100 Factory setting 100 HistoROM AVAIL 831 Display Indicates whether the optional HistoROM M DAT memory module is connected to the electronic insert See also Operating Instructions BA00270P Deltabar S BA00271P Cerabar S or BA00332P Deltapilot S Section 5 5 HistoROM M DAT optional Options Yes HistoROM M DAT is att...

Страница 127: ...ee also this table parameter description for SIM TANK CONT Current see also this table parameter description for SIM CURRENT Note The Flow measuring mode has to be selected in the LINEAR SQROOT parameter to ensure the current output corresponds to the simulated flow value Alarm warning see also this table parameter description for SIM ERROR NO P01 xMx7xxxx 05 xx xx xx 012 Factory setting None SIM ...

Страница 128: ...le column Prerequisite SIMULATION MODE Alarm Warning Factory setting 613 simulation active Table 33 GROUP SELECTION OPERATING MENU DIAGNOSTICS MESSAGES Parameter name Description ALARM STATUS 046 Display Displays the current messages present See also these Operating Instructions Section 8 1 Messages and Section 8 3 Confirming messages On site display The measured value display shows the message wi...

Страница 129: ...he device should behave as in the event of an alarm A or as in the event of a warning W See also ERROR No See also these Operating Instructions Section 8 2 Response of outputs to errors Options Alarm A output current assumes a defined value Warning W device continues measuring On site operation 1 Enter the corresponding message number for ERROR No field 2 Select Alarm or Warning option Digital com...

Страница 130: ...h sensor limit 1 1 For the high sensor limit see PRESS SENS HILIM Tmin ALARM WINDOW 334 Entry Customer specific process monitoring enter lower temperature limit You can use the SELECT ALARMTYPE parameter to enter how the device responds if the operating pressure undershoots the specified value See also these Operating Instructions Section 8 1 Messages table Code E732 and Section 8 2 Response of ou...

Страница 131: ...Measured current 0 2 mA Factory setting 4 mA CURR TRIM 20mA 042 Entry Enter current value for the upper point 20 mA of the current output trim line You can adapt the current output to the transmission conditions with this parameter and CURR TRIM 4mA Perform current trim for the upper point as follows 1 Select SIMULATION group Menu path GROUP SELECTION OPERATING MENU DIAGNOSTICS SIMULATION 2 Select...

Страница 132: ...TATUS Page 128 parameter shows the message with the highest priority See Priority column Note If the device detects a defect in the on site display during initialization special error messages are generated For the error messages see Page 139 Section 8 1 1 On site display error messages For support and further information please contact Endress Hauser Service See also Section 8 4 8 5 and 8 6 Code ...

Страница 133: ...r B factory setting Warning C Out of specification S B Sensor low pressure Pressure too low Increase pressure until message disappears 30 Sensor defect Replace sensor 121 A121 Alarm B Failure F B Checksum error in factory segment of EEPROM Main electronics defect Replace main electronics 5 122 A122 Alarm B Failure F B Sensor not connected Cable connection sensor main electronics disconnected Check...

Страница 134: ...e also these Operating Instructions chapter 4 to 6 Perform reset Code 7864 and carry out calibration again 49 Loose connection at sensor cable Wait a short period of time and tighten the connection or avoid loose connection 700 W700 Warning C Maintenance request M C Last configuration not stored An error occurred when writing or reading configuration data or the power supply was disconnected Perfo...

Страница 135: ...sation table is either below the value for HYDR PRESS MIN or MIN LEVEL or above the value for HYDR PRESS MAX or LEVEL MAX Carry out calibration again See also these Operating Instructions chapter 5 38 710 W710 Warning C Funktion check C B Set span too small Not allowed Values for calibration e g lower range value and upper range value are too close together Adjust calibration to suit sensor See al...

Страница 136: ...evice if necessary 35 Unsuitable download carried out Check configuration and perform download again 719 A719 Alarm B Funktion check C B Y VAL of lin table out of edit limits At least on Y VALUE in the linearisation table is below the MIN TANK CONTANT or above the MAX TANK CONTENT Carry out calibration again See also Operating Instructions BA00274P chapter 5 or these Operating Instructions Page 2 ...

Страница 137: ...ror C Factory setting Warning C Out of specification S C Sensor pressure error overrange Electromagnetic effects are greater than specifications in the technical data See Technical Information TI00382P Deltabar S TI00383P Cerabar S or TI00416P Deltapilot S Block off electromagnetic effects or eliminate source of disturbance 28 Pressure is outside permitted range Check pressure present reduce or in...

Страница 138: ... main electronics 737 A737 Alarm B Failure F B Measurement error Fault in the main electronics Briefly disconnect device from the power supply 20 Main electronics defect Replace main electronics 738 A738 Alarm B Failure F B Measurement error Fault in the main electronics Briefly disconnect device from the power supply 19 Main electronics defect Replace main electronics 739 A739 Alarm B Failure F B...

Страница 139: ... Block off electromagnetic effects or eliminate source of disturbance 11 Main electronics defect Replace main electronics 745 W745 Warning C Maintenance request M C Sensor data unknown Sensor does not suit the device electronic sensor nameplate Device continues measuring Replace sensor with a suitable sensor 56 746 W746 Warning C Funktion check C C Sensor connection error initialising Electromagne...

Страница 140: ...ress Hauser Message Measure Initialization VU Electr Defect A110 Exchange on site display Initialization VU Electr Defect A114 Initialization VU Electr Defect A281 Initialization VU Checksum Err A110 Initialization VU Checksum Err A112 Initialization VU Checksum Err A171 ...

Страница 141: ...arameter The bargraph adopts the value which corresponds to the current value See this table column Alarm or Warning On site display The measured value and message are displayed alternately Measured value display symbol is permanently displayed The measured value and message are displayed alternately Measured value display symbol flashes The measured value and message are displayed alternately Mea...

Страница 142: ... MENU SETTINGS BASIC SETUP In addition there are also parameters that are only displayed if other parameters are appropriately configured For example the Customer Unit P parameter is only displayed if the User unit option was selected for the PRESS ENG UNIT parameter These parameters are indicated with a For a description of the parameters please refer to chapter 7 Description of parameters The ex...

Страница 143: ...value GROUP SELECTION QUICK SETUP LANGUAGE Pressure Level Deltabar S only POS ZERO ADJUST SET LRV SET URV MAX FLOW MAX PRESS FLOW POS ZERO ADJUST POS ZERO ADJUST EMPTY CALIB FULL CALIB POS ZERO ADJUST POS INPUT VALUE CALIB OFFSET DAMPING VALUE DAMPING VALUE DAMPING VALUE SETTINGS OPERATING MENU POSITION ADJUSTMENT There are parameters that are only displayed if other parameters are appropriately c...

Страница 144: ...UE FULL CALIB FULL PRESSURE FULL CALIB EMPTY CALIB Level Standard Level Easy Pressure Level Easy Height Mass MASS FLOW UNIT Gas norm cond Volume p cond NORM FLOW UNIT Gas std cond STD FLOW UNIT UNIT FLOW FLOW MEAS TYPE CUSTOMER UNIT F MAX FLOW CUST UNIT FACT F MAX PRESS FLOW DAMPING VALUE Flow Continuation see the following page Pressure Level SET LRV SET URV GET LRV DAMPING VALUE GET URV BASIC SE...

Страница 145: ...ITY HEIGHT UNIT 100 POINT ZERO POSITION CUSTOMER UNIT H CUST UNIT FACT H DAMPING VALUE EMPTY CALIB FULL PRESSURE EMPTY PRESSURE FULL CALIB ADJUSTED DENSITY Height Linearized COMB MEASURAND LEVEL MIN LEVEL MAX CALIBRATION MODE Wet Dry Height HEIGHT UNIT CUTOMER UNIT H CUST UNIT FACT H Volume UNIT VOLUME CUSTOMER UNIT V CUST UNIT FACT V Mass MASS UNIT CUSTOMER UNIT M CUST UNIT FACT M CUSTOMER UNIT V...

Страница 146: ...ECTION LIN EDIT MODE MEASURING TABLE TANK DESCRIPTION TOTALIZER SETUP RESET TOTALIZER 1 TOTALIZER 2 UNIT TOTALIZER 1 UNIT FACT U U TOTAL 1 NEG FLOW TOT 1 TOT 2 USER UNIT FACT U U TOTAL 2 NEG FLOW TOT 2 TOT 1 USER UNIT LINEARISATION Level measuring mode only Flow measuring mode only Deltabar S only X VAL Y VAL LINE NUMB ACTIVE LIN TAB X ACTIVE LIN TAB Y TANK DESCRIPTION LINEARISATION Level measurin...

Страница 147: ...asuring mode only 2 Display via digital communication 3 Level measuring mode only 5 only LEVEL SELECTION Level Easy Pressure Pressure Level Flow OUTPUT SET MAX ALARM SET MIN CURRENT OUTPUT CURRENT OUTPUT FAIL MODE ALT CURR OUTPUT ASSIGN CURRENT LINEAR SQROOT CURR CHARACT DISPLAY SAFETY CONFIRM ALTERNATE DATA DISPLAY CONTRAST MENU DESCRIPTOR MAIN DATA FORMAT SAFETY LOCK SAFETY PASSWORD CALIB OFFSET...

Страница 148: ...ers are appropriately configured These parameters are indicated with a TRANSMITTER INFO HART DATA MANUFACTOR ID HART MESSAGE BUS ADDRESS DEVICE REVISION PREAMBLE NUMBER HART DATE DEVICE TYPE TRANSMITTER DATA DEVICE SERIAL No CUST TAG NUMBER ELECTR SERIAL No DEVICE DESIGN HARDWARE REV ADDITIONAL INFO SOFTWARE VERSION CONFIG RECORDER PCB TEMPERATURE ALLOWED MIN TEMP ALLOWED MAX TEMP DIP STATUS LONG ...

Страница 149: ...ELECT ALARMTYPE ALARM DELAY ALARM DISPL TIME ALARM STATUS LAST DIAG CODE ACK ALARM MODE ACK ALARM ERROR NO SIMULATION MODE OPERATING HOURS INSERT PIN NO HistoROM AVAIL DOWNLOAD SELECT OPERATION DIAGNOSTICS SIMULATION MESSAGES USER LIMITS Alarm Warning none Level Pressure Tank content Flow Current ENTER RESET CODE There are parameters that are only displayed if other parameters are appropriately co...

Страница 150: ...Appendix Cerabar S Deltabar S Deltapilot S 4 20mA HART 150 Endress Hauser ...

Страница 151: ...URRENT MODE 052 115 CURRENT OUTPUT 875 9 CUST TAG NUMBER 055 117 CUST UNIT FACT F 609 95 CUST UNIT FACT H 705 Height linearized level type84 89 CUST UNIT FACT H 705 Linear level type 71 76 CUST UNIT FACT M 703 Height linearized level type 86 CUST UNIT FACT M 703 Linear level type 73 CUST UNIT FACT M 703 Pressure linearized level type 80 CUST UNIT FACT P 317 56 59 63 68 93 CUST UNIT FACT V 607 Heig...

Страница 152: ...44 LANGUAGE display group 110 LAST DIAG CODE 564 128 LEVEL BEFORE LIN 050 123 LEVEL MAX 712 86 LEVEL MIN 755 86 LEVEL MODE 718 68 LEVEL SELECTION 020 46 49 LIN EDIT MODE 397 on site operation 101 LIN EDIT MODE digital communication 105 LIN MEASURAND 804 70 LINd MEASURAND 805 79 LINEAR SQROOT 390 113 LINEAR SQROOT 854 Safety confirm group 9 LINE NUMB 549 on site operation 102 LINE NUMB digital comm...

Страница 153: ...etup 77 SET LRV 762 Level extended setup 98 SET LRV 852 Safety confirm group 9 SET MAX ALARM 342 113 SET MIN CURRENT 343 113 SET URV 012 Level Easy Height level selection 66 SET URV 012 Level Easy Pressure level selection 61 SET URV 246 Pressure measuring mode 48 57 SET URV 638 Flow extended setup 100 SET URV 720 Level basic setup 77 SET URV 763 Level extended setup 98 SET URV 853 Safety confirm g...

Страница 154: ...tabar S Deltapilot S 4 20mA HART Index 154 Endress Hauser Y Y VAL 551 on site operation 102 Y VAL digital communication 106 Z ZERO POSITION 814 Height linearized level type 90 ZERO POSITION 814 Linear level type 77 ...

Страница 155: ...Cerabar S Deltabar S Deltapilot S 4 20mA HART Index Endress Hauser 155 ...

Страница 156: ...www endress com worldwide BA00274P 00 EN 16 14 71260321 CCS FM SGML 9 71260321 ...

Отзывы: