Description of parameters
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
108
Hauser
TOTALIZER 1 UNIT
(398), (666), (664), (662)
Selection
Select unit for totalizer 1.
Depending on the setting in the FLOW-MEAS. TYPE parameter (
Page 93) this
parameter offers a list of volume, norm volume, standard volume and mass units. When a
new volume or mass unit is selected, totalizer-specific parameters are converted and
displayed with the new unit within a unit group. When the flow mode is changed, the
totalizer value is not converted.
The 3-digit ID number on the on-site display depends on the FLOW-MEAS. TYPE
selected:
– (398): FLOW-MEAS. TYPE "Volume p. cond."
– (662): FLOW-MEAS. TYPE "Mass"
– (664): FLOW-MEAS. TYPE "Gas. std. cond."
– (666): FLOW-MEAS. TYPE "Gas. norm conditions"
Factory setting:
m
3
TOT. 1 USER UNIT (627)
Entry
Enter text (unit) for customer-specific unit for totalizer 1.
You can enter a maximum of eight alphanumeric characters here.
See also FACT. U.
U. TOTAL. 1.
Prerequisite:
• TOTALIZER 1 UNIT = User unit
!
Note!
Only the first five characters are shown on the on-site display. For example, if "crates" is
specified as the customer-specific unit, "crate" is displayed.
If the unit contains a slash, up to eight characters can be shown on the on-site display.
The maximum number of characters in the counter is again limited to five. For example,
if "crates/m2" is specified as the customer-specific unit, "crate/m2" is displayed. In the
FieldCare, all eight characters are displayed.
In the HART handheld terminal, the customer-specific unit is only displayed in the TOT.
1 USER UNIT parameter. The measured value is displayed with the additional text "User
Unit".
Factory setting:
_ _ _ _ _ _ _ _
FACT. U. U. TOTAL. 1 (329)
Entry
Enter conversion factor for a customer-specific unit for totalizer 1.
The conversion factor must be entered in relation to an appropriate SI unit, e.g. m
3
for the
"Volume p. cond." FLOW-MEAS. TYPE.
See also TOT. 1 USER UNIT.
Prerequisite:
• TOTALIZER 1 UNIT = User unit
Example:
You want the measured value to be displayed in "buckets".
– MEASURED VALUE =1 m3
i
100 buckets
– Entry TOT. 1 USER UNIT: bucket
– Entry FACT. U. U. TOTAL. 1: 100
– Result: MEASURED VALUE = 100 buckets
Factory setting:
1.0
NEG. FLOW TOT. 1 (400)
Selection
Specify way of counting negative flows for totalizer 1.
P01-xMD7xxxx-16-xx-xx-xx-003
Factory setting:
Inc. on neg. flow
Table 20: (GROUP SELECTION
)
OPERATING MENU
SETTINGS
TOTALIZER SETUP
Parameter name
Description
Options
Stop on neg. flow
Dec. on neg. flow
Inc. on. neg. flow
negative flow
Total remains constant
Total decreases
Total increases
Total increases
Total increases
Total increases
positive flow