Installation
CUA261
6
Hauser
3.2.2
Product identification
Product page
www.endress.com/cua261
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1. Go to the product page for your product on the Internet.
2. At the bottom of the page, click the link
Online Tools
and then select
Access device
specific information
.
An additional window opens.
3. Enter the order code from the nameplate into the search field and then select
Show
details
.
You will receive information on each feature (selected option) of the order code.
Manufacturer's address
Hauser Conducta GmbH+Co. KG
Dieselstraße 24
D-70839 Gerlingen
3.3
Scope of delivery
The delivery comprises:
• Adapter in the version ordered,
with or without VARIVENT flow assembly N 68 mm
• Clamp connector (only for version with VARIVENT flow assembly)
• Operating Instructions
4
Installation
4.1
Installation conditions
4.1.1
Installation instructions
• Make sure that the optical windows of the flow assembly are vertically aligned. This
prevents buildup on the window surfaces. The optical windows are vertical when the sensor
and the detector housing are in a horizontal position.
• Install the flow assembly and sensor upstream from the pressure regulators.
Содержание CUA261
Страница 20: ...www addresses endress com 71348639 71348639 ...