CNGmass DCI
Troubleshooting
Hauser
63
10.2
System error messages
Serious system errors are
always
recognized by the device as "Fault message", and are shown as a
lightning flash (
$
) on the display! Fault messages immediately affect the outputs. Simulations and
positive zero return, on the other hand, are classed and displayed as "Notice messages".
"
Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer for repair.
Follow the important procedures before returning the flowmeter to Hauser
Always enclose a duly completed "Declaration of Contamination" form. You will find a preprinted
blank of this form at the back of this manual.
!
Note!
• The listed error message types below correspond to the factory setting.
• See also the information on
No.
Error message / Type
Cause
Remedy / spare part
S = System error
$
= Fault message (with an effect on the inputs and outputs)
! = Notice message (without any effect on the inputs and outputs)
No. # 0xx
→
Hardware error
001
S: CRITICAL FAILURE
$
: # 001
Serious device error.
Replace the amplifier board.
Spare parts
011
S: AMP HW EEPROM
$
: # 011
Amplifier:
Defective EEPROM.
Replace the amplifier board.
Spare parts
012
S: AMP SW EEPROM
$
: # 012
Measuring amplifier: Error when accessing data of the
EEPROM.
The EEPROM data blocks in which an error has occurred are
displayed in the "TROUBLESHOOTING" function.
Press Enter to acknowledge the errors in question; default values
are automatically inserted instead of the erroneous parameter
values.
!
Note!
The measuring device has to be restarted if an error has occurred
in a totalizer block
(
È
also error No. 111 / CHECKSUM TOTAL.).
031
S: SENSOR HW DAT
$
: # 031
Sensor DAT:
1.
S-DAT is defective.
2.
S-DAT is not plugged into the amplifier board or is
missing.
1.
Replace the S-DAT.
Spare parts
Check the spare part set number to ensure that the new,
replacement DAT is compatible with the measuring
electronics.
2.
Plug the S-DAT into the amplifier board
71 or
73.
032
S: SENSOR SW DAT
$
: # 032
Sensor DAT:
Error accessing the calibration values stored in the S-
DAT.
1.
Check whether the S-DAT is correctly plugged into the
amplifier board
71 or
2.
Replace the S-DAT if it is defective.
Spare parts
Before replacing the DAT, check that the new, replacement
DAT is compatible with the measuring electronics.
Check the:
– Spare part set number
– Hardware revision code
3.
Replace measuring electronics boards if necessary.
Spare parts
041
S: TRANSM. HW DAT
$
: # 041
Sensor DAT:
1.
T-DAT is defective.
2.
T-DAT is not plugged into the amplifier board or is
missing.
1.
Replace the T-DAT.
Spare parts
Check the spare part set number to ensure that the new,
replacement DAT is compatible with the measuring
electronics.
2.
Plug the T-DAT into the amplifier board
Содержание CNGmass DCI
Страница 2: ......
Страница 86: ...CNGmass DCI Index 86 Endress Hauser ...
Страница 88: ...www endress com worldwide BA138D 06 en 10 09 71096458 FM SGML6 0 ProMoDo ...