Condumax CLS15D/16D/21D
Repair
Hauser
23
Recommended intervals for seal replacement (reference values)
Application
Window
Media with temperatures from 50 to 100 °C (122 to 212 °F)
Approx. 18 months
Media with temperatures < 50 °C (122 °F)
Approx. 36 months
Sterilization cycles, max. 150 °C (302 °F), 45 min.
Approx. 400 cycles
To ensure your sensor is operational again after being exposed to very high loads, you can
have it regenerated in the factory. In the factory, the sensor is fitted with new seals and
recalibrated.
Please contact your sales office for information on replacing the seal and recalibration in the
factory.
8.2
Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure the swift, safe and professional return of the device:
‣
Refer to the website
www.endress.com/support/return-material
for information on the
procedure and conditions for returning devices.
8.3
Disposal
The device contains electronic components. The product must be disposed of as electronic
waste.
‣
Observe the local regulations.