Cleanfit W CPA450
Technical data
Hauser
31
8
Technical data
8.1
Environment
8.2
Process
Ambient temperature
0 to 50 °C (32 to 122 °F)
Process pressure
max. 6 bar (87 psi)
"
Caution!
• The maximum permissible pressure for assembly movement is 2 bar (29 psi)!
• Consider the process conditions of the applied sensor!
Process temperature
-15 to 130 °C (5 to 266 °F)
"
Caution!
Consider the maximum process temperature of the sensor!
Pressure range
max. 16 bar at 40 °C (232 psi at 104 °F), 6 bar at 130 °C (87 psi at 266 °F)
Pressure-temperature load curve
a0010342
Fig. 11:
Pressure-temperature diagram
A
Maximum pressure limit
B
Recommended operating pressure
C
Maximum pemissible pressure for assembly movement
20
0
4
8
12
16
-15
40
60
80
100
120
140
130
A
20
0
4
8
12
16
-15
40
60
80
100
120
140
130
A
B
p [bar]
T [°C]
C
58
116
174
232
p [psi]
2
29
68
32
5
104
140
176
212
248
284
266
T [°F]
6
87