Diagnostics and troubleshooting
Ceraphant PTC31B, PTP31B, PTP33B IO-Link
46
Hauser
10.2.2 Overview of diagnostic events
Status
signal/
Diagnostic
event
Diagnostic
behavior
IO-Link
EventQualifier
EventCode
Event text
Cause
Corrective measure
S140
Warning
IO-Link Warning 0x180F
Sensor signal
outside of permitted
ranges
Overpressure or low
pressure present
Operate device in the specified
measuring range
F270
Fault
IO-Link Error
0x1800
Overpressure/low
pressure
Overpressure or low
pressure present
• Check the process pressure
• Check sensor range
• Restart device
F270
Fault
IO-Link Error
0x1800
Defect in
electronics/sensor
Defect in electronics/
sensor
Replace device
C431
Warning
IO-Link Warning 0x1805
Invalid position
adjustment (Current
output)
The adjustment performed
would cause the sensor
nominal range to be
exceeded or undershot.
Position adju parameter of
the current output must be within
the sensor nominal range
• Check position adjustment (see
Zero point configuration (ZRO)
parameter)
• Check measuring range (see
Value
for 20 mA (STU)
and
Value for 4
mA (STL)
parameters)
C432
Warning
IO-Link Warning 0x1806
Invalid position
adjustment
(Switching Output
1)
The adjustment performed
causes the switch points to
be outside the sensor
nominal range.
Position adju parameter of
the hysteresis and window function
must be within the sensor nominal
range
• Check position adjustment (see
Zero point configuration (ZRO)
parameter)
• Check the switch point,
switchback point for hysteresis
and window function
C432
Warning
IO-Link Warning 0x1807
Invalid position
adjustment
(Switching Output
2)
The adjustment performed
causes the switch points to
be outside the sensor
nominal range.
Position adju parameter of
the hysteresis and window function
must be within the sensor nominal
range
• Check position adjustment (see
Zero point configuration (ZRO)
parameter)
• Check the switch point,
switchback point for hysteresis
and window function
F437
Fault
IO-Link Error
0x1810
Incompatible
configuration
Invalid device
configuration
• Restart device
• Reset device
• Replace device
C469
Fault
IO-Link Error
0x1803
Switch points for
output 1 violated
Switch point ≤ switchback
point
Check switch points at output
C469
Fault
IO-Link Error
0x1809
Switch points for
output 2 violated
Switch point ≤ switchback
point
Check switch points at output
C485
Warning
IO-Link Warning 0x8C01
Simulation active
During simulation of the
switch output or current
output, the device issues a
warning message.
Switch off simulation.
S510
Fault
IO-Link Error
0x1802
Turn down violated
A change in the span
results in a violation of the
turn down (max. TD 5:1)
Values for adjustment
(lower range value and
upper range value) are too
close together
• Operate device in the specified
measuring range
• Check the measuring range