Description of parameters
Cerabar S 4 to 20mA HART
30
Hauser
PRESSURE (301)
Display
Displays the measured pressure after sensor recalibration, position adjustment and
damping. This value corresponds to the MEASURED VALUE parameter in the "Pressure"
measuring mode.
P01-MIDxxxxx-05-xx-xx-EN-001
CORRECTED PRESS. (434)
Display
Displays the measured pressure after sensor trim and position adjustment and before
damping.
See also PRESSURE diagram.
SENSOR PRESSURE (584)
Display
Displays the measured pressure before sensor trim, position adjustment and damping.
See also PRESSURE diagram.
SENSOR TEMP. (367)
Display
Displays the temperature currently measured in the sensor. This can deviate from the
process temperature.
MEAS. VAL. TREND (378)
Display
Displays the trend of the pressure measured value.
Possibilities: increasing, decreasing, constant
Table 12: (GROUP SELECTION
) OPERATING MENU
PROCESSINFO
PROCESS VALUES "Pressure"
Parameter name
Description
P
I
Sensor
Current
output
Pressure
PRESSURE
Damping
CORRECTED
PRESS.
Position
adjust-
ment
Sensor
trim
SENSOR
PRESSURE