CCS120D
Maintenance
Hauser
31
3. Rinse electrodes thoroughly with clean, warm water ensuring all residue of electrolyte is
removed.
4. Allow the electrodes to dry.
5. Screw a new membrane cap loosely onto the electrodes to protect them.
6. Fit protection cap on sensor → 31.
7. For recommissioning, follow the same procedure as for commissioning → 24.
Ensure that no biofouling occurs during longer interruptions to measurement. Remove
continuous organic deposits such as films of bacteria from media with a high
concentration of chlorine.
Fitting the protection cap on the sensor
1. To keep the membrane moist after the sensor has been removed, fill the protection cap
with clean water.
A0037886
13
Carefully slide protection cap onto the membrane cap.
2. Top part of protection cap is in the open position.
Carefully slide protection cap onto the membrane cap.