Installation and commissioning
multi N/C pharma HT
54
4.3.3
EPA Sampler
CAUTION
Risk of injury from moving parts
There is a risk of injury in the movement range of the sampler arm. For example, hands
or fingers might be crushed.
¡
Maintain a safety distance from the sampler during operation.
NOTICE
Risk of device damage
If the sampler arm is obstructed during operation, the drives can be destroyed.
¡
Do not touch the sampler arm during operation.
¡
Only carry out manual adjustment when the device is switched off.
The autosampler has a piercing function for sample vessels with septum caps. The sam-
pler can be equipped with 1 to 2 canulas.
Technical data
Maximal sample number
64
Sample vessels
40 ml
Operating voltage
24 V DC via external power supply
Power consumption
30 VA
Grid voltage of external power supply
100 to 240 V, 50/60 Hz (autosensing)
Dimensions (W x H x D)
500 x 560 x 400 mm
Layout
Figure 37EPA Sampler autosampler
1 Connection hoses to the analyzer
2 Sample tray
3 Wash cup
4 Holding-down clamp
5 Special canula
6 Autosampler arm with canula holder