multi N/C pharma HT
Function and design
21
The top middle connection on the glass container is connected to the phosphoric acid
pump. The phosphoric acid pump pumps phosphoric acid (10 %) into the TIC reactor for
each TIC determination.
The cooling block dries the measuring gas by freezing out the water vapor. The dry mea-
suring gas is routed via the top connection out of the TIC condensate container. The
measuring gas drying is maintenance-free.
The condensate pump pumps the condensate or waste solution from the TIC determina-
tion out via the bottom side outlet on the glass container after each measurement
Figure 12TIC condensation module
1 TIC condensate container
2 Connection for hose 1/condensation coil
measuring gas feed
3 Cooling block
4 Connection to the water traps
5 Water traps
6 Connection for hose AD/phosphoric acid
feed
7 Connection for hose BB/direct sample
feed for TIC detection
8 Connection to condensate pump (waste
hose 11)
Water traps
The water traps remove interfering components from the measuring gas and protect the
detector and the gas box. The water traps are mounted in the gas path behind the cool-
ing block or behind the gas box. The water traps each consist of a larger and a smaller
water trap. The larger water trap (TC prefilter) retains aerosols during operation. The
smaller water trap (disposable retention filter) retains rising water.