multi X 2500
AOX determination
71
2.
Insert the storage container for the combined electrode into
the bracket in the door of the chlorine module.
3.
Insert the combined electrode into the opening of the
measuring cell.
4.
Insert the measuring gas hose from the sulfuric acid
container using the Teflon connector (1) into the connection
of the combined electrode.
5.
Connect the olive-tip suction adapter to the exhaust air hose
in the chlorine module.
6.
Connecting the combined electrode and the measuring cell at
the inside of the rear panel of the chlorine module:
Connection combined electrode (1)
Connection measuring cell (2)
6.3.2
Preparing the measuring cell
The preparation of the measuring cell for daily measurements includes the following
steps:
Preparation of the electrolyte solution
Execution of the end point routine
See chapter "Preparing the measuring cell", p. 59 for information on how to prepare
the measuring cells "sensitive" and "high concentration".