multi X 2500
Fault removal
113
11.2.3
Analytical problems
Combustion at the cannula
Cause
Solution
Argon and oxygen connection on the multi-purpose
combustion tube were mixed up
Connect gases correctly (1-2 in Fig. 10, p. 23)
Low results independent of detection
Cause
Solution
Metering fault
System leaking
Temperature set too low
Wrong or unsuitable calibration
Sample loss at or before sample introduction due to
vaporization or spillage
Insufficient post-combustion period
Sooting in the system
Check metering
Check system tightness (
→
Check temperature configurations in the method
Check calibration, repeat if required
Close liquid samples
Check the sampler routine for solid samples
Particularly for solids, set a post-combustion time of
120 s in the method
Clean or replace sooted parts
Carryover
Cause
Solution
Inadequate rinsing of the sample introduction
components
Combustion tube insufficiently rinsed for EOX
determinations
Contamination within the injection port of the
combustion tube or within the ABD sluice
Rinse metering syringes prior to sampling
Rinse combustion tube with clean solvent, i.e. blank
measurements until values are constant
Replace septum, clean sluice
Scattering measured values
Cause
Solution
Metering faulty
Combustion tube contaminated or severely crystallized
Check metering
Clean or replace combustion tube