multi N/C 2100S (duo, pharma)
Installation and commissioning
37
4
Installation and commissioning
4.1
Installation conditions
4.1.1
Ambient conditions
¡
This laboratory device is designed for inside use.
¡
Avoid direct sunlight and radiation from heaters onto the device. If necessary, pro-
vide air conditioning.
¡
The installation site must be free of drafts, dust and caustic fumes.
¡
The room air must be as low in TOC and NO
x
as possible.
¡
Avoid mechanical shocks and vibrations.
¡
Do not locate the device near sources of electromagnetic interference.
¡
Place the device on a heat-resistant and acid-resistant surface.
¡
The device must be positioned in such a way that allows easy access from all sides.
¡
Keep the ventilation slits free and do not obstruct them with other devices.
The following climate requirements apply in the room of operation:
Operating temperature
+10 to 35 °C (air-conditioning recom-
mended)
Maximum humidity
90 % at 30 °C
Air pressure
0.7 to 1.06 bar
Storage temperature
5 to 55 °C
Humidity during storage
10 to 30 % (use desiccant)
Operating altitude (max.)
2000 m
4.1.2
Device layout and space requirements
The basic device has been designed as a compact bench-top device. The required space
depends on all the components that make up the measuring station.
The liquid autosampler is mounted on the top of the basic device. The required height
results from the height of the basic device and the autosampler. There must be a clear-
ance of at least 10 cm between the device system and any cabinet/shelf located above
it.
Further components of the measuring station:
¡
The PC, monitor and printer may be placed on a separate side table.
¡
An acid-resistant waste container can be placed on or under the bench.
¡
The CLD nitrogen detector is set up to the right of the basic device.
Optional modules:
¡
The HT 1300 and the manual TIC solids modules are placed to the right of the basic
device.
¡
The Double Furnace Module is attached to the left side panel of the basic device.
¡
The ChD (approx. 0.5 kg) nitrogen detector is installed in the basic device.
Component
Dimensions (width x height x
depth)
Weight
Basic device
513 x 464 x 550 mm
30 kg
AS 60 autosampler
510 x 500 x 400 mm
9 kg
CLD nitrogen detector
300 x 470 x 550 mm
12.5 kg