Function and design
multi N/C 2100S (duo, pharma)
22
The condensate pump pumps the condensate or waste solution from the TIC determina-
tion out via the bottom side outlet on the glass container after each measurement
The condensation coil and the TIC condensation module are only used for drying and
cleaning measuring gas in the multi N/C 2100S pharma model. TIC is not determined
here.
Figure 12Condensation coil and TIC condensation module
1 Condensation coil
2 Cooling block
3 TIC condensate container
Water traps
The water traps remove interfering components from the measuring gas and protect the
detector and the gas box. The water traps are mounted in the gas path behind the cool-
ing block or behind the gas box. The water traps each consist of a larger and a smaller
water trap. The larger water trap (TC prefilter) retains aerosols during operation. The
smaller water trap (disposable retention filter) retains rising water.