14.13.2 Replacing the electrolyte solution
1
2
3
5
6
7
8
4
Figure 75Coulometric sulfur detector with measuring cell (without door)
1 Indicator electrodes connection
2 Indicator electrode (black)
3 Gas inlet
4 Measuring cell
5 Port for manual dosing
6 Anode (yellow)
7 Cathode (red)
8 Generator electrodes connection
ð
Replace the electrolyte solution daily and when it is used up.
}
Switch of the magnetic stirrer via the rotary switch.
}
Switch off the detection module via the power switch.
}
Disconnect the two electrode cables on the "generation" and "indication" connections.
}
Disconnect hose 72 from the HX absorber.
}
Remove the measuring cell from the detection module.
}
Remove the electrodes and the gas inlet tube from the measuring cell. Empty the re-
maining electrolyte solution from the generator electrodes. Set the components
down.
}
Remove the electrolyte solution from the measuring cell.
}
Rinse the stirring rod with ultrapure water. Rinse out the measuring cell.