Biometra TAdvanced
Creating, editing and saving programs
Version 02_20
47
Select B
LANK
whenever a new program should be created and select another template
from the list whenever an existing program should be edited.
The program structure can be controlled in the program preview (see Fig. 26). If a
template is selected, the program steps and some additional information are displayed
in spreadsheet form. By using the program preview function, templates can be checked
before they are opened.
5.
To open a program template, press the [O
PEN TEMPLATE
] button.
O
PEN TEMPLATE
Fig. 26 Program template screen
7.3
Assigning program names
Each template has a specific name which is superimposed in the programming screen.
To enter a name, touch the N
AME
input field and enter a new name using the Biometra
TAdvanced keyboard (see section "Keyboard" p. 26).
Note
: If a character is pressed, the pre-displayed name will be deleted and a new
program name can be set. If the program name should be modified, please first press
another button like delete or space.
7.4
Setting the heated lid temperature
To set the heated lid temperature, touch the L
ID
input field. A small window opens in
the middle of the screen with a keyboard (see Fig. 27). Enter a value between 30 and
110 using the Biometra TAdvanced keyboard.
Note
: The default setting for the heated lid temperature is 99 °C. 99 °C is sufficient for
typical PCR programs.
Use the corresponding buttons to switch the heated lid on or off.