ENDEVCO MODEL 136 DC AMPLIFIER
INSTRUCTION MANUAL
IM136
Page 16
2.4.11
Monitoring State
This function selection provides for programming the desired state of the amplifier while in the
normal mode of operation (described in 2.2 and 2.3 above). The following messages will be
shown sequentially on the LED display each time the up-down EDIT keys are pressed: (1)
“EU”, (2) “Vout” and (3) “OFF”.
1.
"EU": RMS voltage at Vout is monitored and shown on the LED display in EU at the
rate of approximately two (2) times per second. The values displayed on the LED
display range from 0.000 to 9999. The value shown on the LED display is computed as
follows:
Display (EU) =
VRMS at Output
Output Scaling
2.
"Vout": RMS voltage at Vout is monitored and shown on the LED display in VRMS at
the rate of approximately two (2) times per second. The values displayed on the LED
display range from 0.00 to 9.99. The value shown on the LED display is computed as
follows:
Display (VRMS) = VRMS at Output
3.
"OFF": The LED display will be turned OFF. This setting provides the lowest noise
performance.
The “Function LED” corresponding to the Monitor function will remain OFF during the normal
mode of operation if the user selected the “OFF” setting. The function LED will remain ON if
monitoring was enabled by selecting any of the other settings listed in the table above.